SPEEDWOLF OPTICAL POWER METER USER’S MANUAL PM630 SHENZHEN SPEEDWOLF TECHNOLOGY CO.,LTD (V2018.1.
SPEEDWOLF—Optical Power Meter—PM630 User’s Manual 1. Model Introductions PM630-A: Measurement range: -70~+10dBm,calibrated wavelength:850nm、1300nm、1310nm、1490nm、1550nm、 1625nm PM630-C: Measurement range: -50~+26dBm,calibrated wavelength:850nm、1300nm、1310nm、1490nm、1550nm、 1625nm 2.
SPEEDWOLF—Optical Power Meter—PM630 User’s Manual under absolute test mode 3. Unit:mW、uW、nW、dBm、dB. 4. Battery level : Indicate the current battery level. When the icon is blank, it means the power is too low, please change the batteries. 5. Auto-off :Activate the 10-minute auto off function 6. ON/OFF 7. Wavelength select 8. VFL:VFL status 9. DBm/w: Absolute power 10. Optical detector 11. Visual Laser(650nm) 12. VFL display 5. Function Keys 1) ON/OFF:press ON/OFF to turn it on.
SPEEDWOLF—Optical Power Meter—PM630 User’s Manual this time the four row units are all displayed). Press λ key to choose the wavelength for calibration. Now connect it with a pigtail which we already know its power and press REF or dBm/W to increase or decrease the power value by 0.05d every time. Until the LCD display the pigtail’s power (Or the difference gap less than 0.3db). Short press power key to save calibration data (backlight will flash once).
SPEEDWOLF—Optical Power Meter—PM630 User’s Manual 2) Fail to provide the warranty card 3) Alter and omit the warranty card 4) Wrong operation including the problem caused by human actors, abnormal working environment and so on. 5) 6) Problems or damages are not caused by the products quality Do not operate it according to this manual. 3. For the following situation, we do not repair it 1) Damaged seal label.