6.BATTERY SAVER (SLEEP MODE)
The meter will automatically enter the sleep mode if there is no
function change or button press for 10minutes, but it works as soon
as you turn the rotary function switch or push down any button
.
7.POWER TOOLS AND AMALL APPLIANCES
This test would also apply to other similar equipment that has a line
cord. For double insulated power tools, the megohmmeter lead
shown connected to the housing would be connected to some metal
part lf the tool(e..g chuck,blade).
Note: The switch of the device must be in the “ON” position and the
main power should be disconnected.
MOTORS
AC-Disconnect the motor from the line by disconnecting the wires at
the motor terminals or by opening the main switch. If the main switch
is used and the motor also has a starter then the starter must be
held, by some means, in the “ON” position. In the latter case, the
measured resistance will include the resistance of the motor, wire
and all other components between the motor and the main switch. If
a weakness is in dicated, the motor and other components should
be checked individually. If the motor is disconnected at the motor
terminals, connect one megohmmeter lead to the grounded motor
housing and the other lead to One of the motor leads.
DC-Disconnect the motor from the line. To test the brush rigging,
field coils and armature connect one megohmmeter lead to the
grounded motor housing and the other lead to the brush on the
commutator. If the resistance measurement indicates a weakness,
raise the brushes off the commutator and separately test the