3. OPERATING INSTRUCTIONS
3-A. MEASURING PROCEDURES - See section 2, page 3 for button
locations and descriptions.
Connect the BNC PLUG (4) to the BNC SOCKET (5).
Slide the PROBE HANDLE (3) into the PROBE HOLDER (2) (opt).
Press the POWER (7) button to turn the unit on.
The default mode is: Auto range, “A” frequency weighting and Fast
time weighting.
To change the frequency weighting, press the A/C (11) button.
The "A" frequency weighting simulates human ear response. For
an environmental sound measurement, select the "A" weighting.
The "C" weighting approximates a flat response. Often, "C" is
used to check the noise of machinery where the target sound level
is already known.
For more information about Frequency Weighting Characteristics,
see the chart on page 8.
Point the MICROPHONE (1) at the sound source.
Use the auto range setting (default) or find the appropriate range
using the RANGE (13) button.
The manual selections are: 30~80dB, 50~100db, or 80~130db.
If “- - - - -” is displayed, switch to another range. The sound level is
displayed in decibels (dB).
For general applications, press the FAST/SLOW (12) button to
select "Fast", simulating the human ear's response time.
The "Slow" setting is generally used to obtain an average of
vibrating sound levels.
Press the POWER (7) button to turn the unit off.
3-B. HOLD
During the measuring procedure, press the HOLD (8) button to
freeze the display.
"HOLD" and the last measurement are displayed.
Press the HOLD (8) button again to exit this function.
3-C RECORD MAXIMUM / MINIMUM MEASURMENTS
Press the REC MAX/MIN (9) button once to enter the recording
mode. "REC" and the recorded values are displayed and
continuously updated.
Press the REC MAX/MIN (9) button as needed to view the
recorded Max and Min values. "REC Max" and the maximum
recorded value, or "REC Min" and the minimum recorded value
are displayed.