User Manual

- 5 -
Note: The Max/Min values are frozen and not updated until the meter
returns to recording mode.
To return to recording mode, press the HOLD (8) button. "REC" is
displayed without "Max" or "Min".
To erase the recorded Max/Min values and exit the recording mode,
press the REC MAX/MIN (9) button for at least 2 seconds.
3-D. MAX HOLD: (Holds the max measurement on the display.)
Press the MAX HOLD (10) button, “P.H” appears on the LCD.
Press the MAX HOLD (10) button again to exit this function.
NOTE: Select auto range when using this function in slow varying
noise environments. Select the appropriate manual range when
using this function in pulsing noise environments.
3-E. AC OUTPUT
The 3.5 mm diameter AC OUTPUT (14) may be used to connect the unit
to an external output device, such as an analyzer, recorder, or controller.
3-F. CALIBRATION
The meter's CAL (15) adjustment is located on the side of the unit.
Use Sper Scientific's 2 Pt. Acoustical Calibrator 850016 (or
equivalent) to calibrate the unit in compliance with OSHA.
Press the POWER (7) button to turn the unit on.
Turn the acoustical calibrator on and place it onto the meter's
MICROPHONE (1)
To calibrate the meter to 94 dB, select the 50 ~ 100 dB range,
FAST response, and "A" weighting.
Use the calibration screw driver to gently turn the CAL (15)
adjustment until the display reads 94.0.
To calibrate the meter to 114 dB, select the 80 ~130 dB range,
FAST response, and "A" weighting and turn the CAL (15)
adjustment until the display reads 114 dB.
3-G. PRECAUTIONS
Do not store or operate the unit in high temperatures or in high
humidity for long periods.
Keep the microphone dry.
Do not force the CAL (15) adjustment, doing so may damage the
mechanism and void the warranty.