Conductivity and TDS Probe 850084 Instruction Manual SPER SCIENTIFIC LTD.
TABLE OF CONTENTS I. INTRODUCTION ............................................................................. 2 II. PANEL DESCRIPTION ................................................................... 2 III. MEASUREMENT PROCEDURES .................................................. 3 IV. CALIBRATION ................................................................................ 3 V. SPECIFICATIONS .......................................................................... 4 VI. WARRANTY.................
III. MEASUREMENT PROCEDURES • Plug probe into the PROBE INPUT SOCKET. • Make sure the PROBE LOCK is in the lock position. • Turn on the instrument by pressing the POWER button. • Press the C & DOWN (CD/TDS) button to select the mode. The meter defaults to conductivity. • The default ATC factor is 2% per °C, adjustable between 0 to 5.0%. To adjust the ATC factor, press the D (TEMP.C) button once, then use the UP, LEFT, and DOWN buttons to set the desired ATC factor. Press the D (TEMP.
V. SPECIFICATIONS Unit of Msrmt. Range Resolution mS 0.2 ~ 2.000 0.001 mS 2 ~ 20.00 0.01 PPM 132 ~ 1320 1 PPM 1320~13,200 10 ºC 0 ~ 60 ºF 32 ~ 140 Accuracy ±(3%fs + 1d) Full Scale 0.1 Operating Temperature (meter) 0°C to 50°C (32°F to 122°F). Operating Humidity Max. 80% RH. Probe Size Diameter: " (22mm). Length: " (120mm). 0.8 1.5 Refer to the Water Quality Meter 850081 Instruction Manual for meter specifications.