Instruction Manual

18
SETUP MODE
View Slope & Offset (pH Probe)
1. Press MODE to select the probe type as pH.
2. Press SET for 2 seconds to enter setup.
3. Press until “ELE” appears in the middle display and
P3.0 appears in the lower display.
4. Press
to enter P3.1, the LCD displays one of
four available slope values; P3.1, P3.2, P3.3, P3.4. If
the value is less than 75% or more than 115%, change
the probe immediately.
5. Press to enter P3.2, P3.3, and P3.4.
Note...
The solution range differs between NIST and Custom
buffers.