LeadAdvancedAlongDorsalAspectofDRG
WARNING:Do not use excessive force to push the lead or sheath into the neural
foramenasthismayresultinpermanentortransientnervedamage.Thepatientshould
beawakeandconversantduringthispartoftheproceduresotheycanprovidefeedback
tothephysician.
IntraoperativeTesting
1. Connecttheheadtothecable.Pressandholdthecablebuttondowntoreleasethe
lockingmechanismandslidetheproximal endofthe leadsintothehead.Releasethe
cablebuttontolocktheleadintoplace.Verifythattheleadcomestoastopbefore
releasingthebutton.Thiswillensurethattheelectricalcontactsareintheappropriate
position.
2. PuttheTNSinstandbymodeorturntheamplitudeofeachleadsettozero.
PRECAUTION:Put the trial stimulator in standby mode or reduce amplitude of leads to
zero before plugging in the cable.Failure to do so may result in delivering an
uncomfortablestimulationtothepatient.
3. PasstheproximalendoftheConnectorCableoffthesterilefieldandconnectto
theTNS.