User's Manual

Merlin™@home Transmitter Model EX1150 User’s Manual 9
Telephone accessories
Telephone accessories included with the transmitter are:
Telephone cable
Phone line splitter
An optional telephone accessory is the Model EX 1151 USB cellular adapter.
Electromagnetic Compatibility
The Merlin™@home transmitter requires special precautions with regard to electromagnetic compatibil-
ity (EMC) and should be used in accordance with the information provided in this manual.
The Merlin@home transmitter complies with the requirements of the international EMC standard
IEC 60601-1-2:2001 when used with the cables listed in ”Telephone accessories” (page 9).
The Merlin@home transmitter is intended for use in the electromagnetic environment specified in
Tables 5 through 8. The user should ensure that it is used in such an environment.
WARNING
The Merlin@home transmitter complies with the limits for
medical devices contained in IEC/EN 60601-1-2:2001.
However, the Merlin@home transmitter may cause radio
interference or may disrupt the operation of nearby
equipment. It may be necessary to mitigate this effect by
reorienting or relocating the receiving device or shielding the
location.
Test Compliance Electromagnetic Environment – Guidance
RF Emission
CISPR 11
Group 1 The Merlin™@home transmitter must emit
electromagnetic energy in order to perform their
intended function. Nearby electronic equipment may
be affected.
RF Emission
CISPR 11
Class B The Merlin@home transmitter is suitable for use in
all establishments, including domestic
establishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class B
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Complies
Table 5. Guidance and manufacturer’s declaration – electromagnetic emissions