User's Manual

10 Electromagnetic Compatibility
Test IEC 60601 Test Level (Actual Level)
1
1. Figures in parentheses are the immunity compliance levels for each test.
Electromagnetic
Environment –
Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±6 kV contact
(±6 kV contact)
±8 kV air
(±8 kV air)
None
Electrical fast transient/burst
IEC 61000-4-4
±1 kV for input/output lines
(±1 kV for input/output lines)
None
Power Frequency (50/60 Hz)
Magnetic Field
IEC 61000-4-8
3 A/m
(3 A/m)
None
Table 6. Guidance and manufacturer’s declaration – electromagnetic immunity