* maximum deviation occurs under unfavourable conditions such as bright sunlight or
when measuring to poorly reflecting or very rough surfaces. Measuring accuracy
between 10 m and 30 m may deteriorate to approx. ± 0.025 mm/m, for distances
above 30 m to ± 0.1 mm/m. In long range mode the maximum deviation from a
distance of 30 m increases to +/- 0.15 mm/m.
Measuring conditions
Measuring range
The range is limited to 200 m.
At night or dusk and if the target is in shadow the measuring range without
target plate is increased. Use a target plate to increase the measurement
range during daylight or if the target has poor reflection properties.
Target surfaces
Measuring errors can occur when measuring toward colourless liquids (e.g.
water) or dust free glass, Styrofoam or similar semi-permeable surfaces.
Aiming at high gloss surfaces may deflect the laser beam and lead to measu-
rement errors.
Against non-reflective and dark surfaces the measuring time may increase.
Distance measurements:
Measuring accuracy up to 10 m
(2
σ
, standard deviation)
typically: ± 1.0 mm*
Range (use target plate from about 100 m) 0.05 m to 200 m
Smallest unit displayed 0.1 mm
Distance measurement
9
Minimum/maximum measurement, Conti-
nuous measurement
9
Area/volume calculation of room data
9
Addition / subtraction
9
Indirect measurement using
Pythagoras
9
Trapezium measurement
9
Tilt measurements:
Tilt sensor:
Accuracy
(2
σ
, standard deviation)
- to laser beam
- to the housing
± 0.3°
± 0.3°
Indirect measurement using tilt sensor
(direct horizontal distance)
9
Angle measurement using tilt sensor
(
± 45°)
9
General:
Laser class II
Laser type 635 nm, < 1 mW
Ø laser point
(at distances)
6 / 30 / 60 mm
(10 / 50 / 100 m)
Autom. laser switch-off after 3 min
Autom. instrument switch-off after 6 min
Display illumination
9
Multifunctional endpiece
9
Timer (self-triggering)
9
Save constant value
9
Historical storage 20 values
Tripod thread (Type: 1/4-20)
9
Battery life,
Type AA, 2 x 1.5V
up to
5 000 measurements
Protection against splashes and dust IP 54, dust-proof,
splash-proof
Dimensions 143.5 x 55 x 30 mm
Weight (with batteries) 195 g
Temperature range:
Storage
Operation
-25°C up to +70°C
(-13°F up to +158°F)
-10°C up to +50°C