User's Manual

STONEX R2 PLUS
116
NOTE:
1. When adjusting the screws of reticle, firstly loosen the screw on the moving direction
of reticle, secondly tighten another screw by the same mount, clockwise turning is for
tightening, and anticlockwise turning is for loosening, the turning mount for tightening or
loosening should be same.
2. After the reticle adjustment, it is necessary to adjust the vertical index error by program,
see “§23.10 Vertical Index Error”.
26.9 Tilt Sensor
If the tilt angle shown on the display shifts from tilt angle 0°(zero point), the instrument
is not correctly leveled. This will adversely affect angle measurement. Perform the
following procedures to cancel the tilt zero point error.
. Check
Set-up the instrument on tripod or adjustment platform, and level it accurately.
1. Press F3: [
0SET] twice to set HA to 0°on P1 in
measure mode.
Press key {
Cnfg} to enter into config mode.
2. Select “
3.Inst adjust to set instrument basic
parameters, two items display.
3. Select “
1. Tilt correct”.
4. Wait a few seconds for this display to stabilize, then
read the automatically compensated angles X1 and
Y1.
Config
1.Meas condition
2.Inst config
3.Inst adjust
4.Com setting
5.Unit
6.Date & Time
Inst adjust
1.Tilt correct
2.Index error correct
Tilt offset
X -0.0020gon
Y 0.0032gon
HA 127.1202gon
Take F1 I
0OK0