Testing and
Calibrating the
UV Lamps
TheȱeffectiveȱUVȱradiationȱfromȱtheȱlampsȱusedȱforȱcuringȱmodelsȱcanȱ
changeȱoverȱtime.ȱToȱensureȱoptimumȱcuringȱofȱmodelsȱduringȱprinting,ȱaȱ
popȬupȱmessageȱremindsȱyouȱtoȱtestȱtheȱlampsȱandȱcalibrateȱtheirȱeffectiveȱ
UVȱradiationȱafterȱeveryȱ300ȱhoursȱofȱprinting.ȱYouȱdoȱthisȱbyȱrunningȱthe
ȱ
UVȱCalibrationȱWizard.
ToȱmeasureȱtheȱUVȱradiation,ȱyouȱneed:
StratasysȱTOLȬ03005ȬSȱ(UVȱsensorȱandȱcable,ȱsuppliedȱwithȱtheȱprinter)
or—
aȱstandȬaloneȱUVȱradiationȱmeter,ȱapprovedȱforȱuseȱwithȱtheȱprinter
Theȱwizardȱcomparesȱtheȱmeasuredȱradiationȱtoȱtheȱrecommendedȱ
radiationȱlevelȱforȱeachȱofȱtheȱlamps,ȱatȱeachȱprintingȱmode—HighȱSpeed,ȱȱ
HighȱQualityȱandȱDigitalȱMaterial.
Whenȱcalibratingȱtheȱlamps,ȱtheȱwizardȱattemptsȱtoȱadjustȱtheȱradiationȱ
level,ȱifȱnecessary.
Ifȱtheȱreadingȱisȱwithinȱtheȱacceptableȱrange,ȱtheȱwizardȱcontinuesȱtoȱ
theȱnextȱphase.ȱ
Ifȱfurtherȱadjustmentȱisȱnecessary,ȱtheȱcurrentȱphaseȱisȱrepeated.ȱ
IfȱtheȱUVȱradiationȱisȱtooȱlowȱtoȱbeȱproperlyȱadjusted,ȱtheȱwizardȱ
continuesȱtoȱtheȱnextȱphase,ȱbutȱtheȱfinalȱwizardȱscreenȱindicatesȱthatȱ
theȱlamp’sȱradiationȱforȱtheȱprintingȱmodeȱisȱunacceptable.
To test and calibrate UV lamp radiation:
1. StartȱtheȱUVȱCalibrationȱWizardȱfromȱtheȱOptionsȱmenuȱ(seeȱfigure 6Ȭ22ȱ
onȱpage 6Ȭ23).
2. Inȱtheȱfollowingȱscreen,ȱselectȱ
Calibrate UV Intensity toȱadjustȱtheȱlamp’sȱ
radiationȱlevelȱtoȱtheȱacceptableȱrange.ȱ
Figure 6-82: UV Test/Calibration selection
Before testing and calibrating the UV lamps:
¾ Make sure the glass lens on each of the UV lamps is clean. For easy
access, run the Head Cleaning Wizard, and use the special cleaning pad
supplied.
¾ Make sure that the glass over the UV radiation sensor is clean.