After examining the UV calibration results:
1. ClickȱNext.
2. RemoveȱtheȱUVȱsensorȱandȱcloseȱtheȱprinterȱcover.ȱAfterȱconfirmingȱ
theȱitemsȱlistedȱinȱtheȱwizardȱscreen,ȱclickȱ
Next.
Figure 6-94: Final UV Calibration Wizard screen
Calibrating the
Load Cells
Loadȱcellsȱareȱsensorsȱthatȱmeasureȱtheȱweightȱofȱtheȱmaterialȱcartridgesȱ
andȱtheȱwasteȱcontainerȱinȱtheȱprinter.ȱItȱisȱimportantȱthatȱyouȱperiodicallyȱ
checkȱthatȱtheȱweightȱmeasurementsȱareȱaccurate,ȱbothȱforȱconvenienceȱandȱ
toȱpreventȱunnecessaryȱwasteȱofȱprintingȱmaterials.ȱItȱisȱrecommendedȱthatȱ
youȱcalibrateȱtheȱload
ȱcellsȱonceȱaȱmonth.
To calibrate load cells:
1. StartȱtheȱLoadȱCellȱCalibrationȱwizardȱfromȱtheȱOptionsȱmenuȱ(seeȱ
figure 6Ȭ8ȱonȱpage 11).ȱ
2. Inȱtheȱfollowingȱscreen,ȱselectȱoneȱorȱmoreȱloadȱcellsȱthatȱyouȱwantȱtoȱ
calibrate.ȱ
Figure 6-95: Load cell selection
3. Removeȱeachȱselectedȱcartridgeȱfromȱtheȱmaterialsȱcart,ȱandȱclickȱNext.