User's Manual

17
DRAFT
To Program Footswitch Options:
From the Adjust Options screen, touch the Program Footswitch button to program the footswitch pedals and pads based
on the capabilities of the connected handpiece. Two footswitches may be connected to and simultaneously operated
from the same console. However, two footswitches cannot be used to control the same handpiece at the same time.
Assign Specific Option
Select a footswitch pedal or pad [A, B, I, II
or III] button and assign a specific option.
View Handpiece Port
Indicates the handpiece port that the
footswitch is programmed for.
Based on handpiece capabilities, options may include:
FORWARD: pressure on the pad or pedal will cause the
handpiece to rotate in the forward (clockwise) direction.
REVERSE: pressure on the pad or pedal will cause the
handpiece to rotate in the reverse (counterclockwise) direction.
+OSC TURNS: pressure on the footswitch pad will cause the
handpiece to oscillate; increase number of turns.
-OSC TURNS: pressure on the footswitch pad will cause the
handpiece to oscillate; decrease number of turns.
•ON: pressure on the footswitch pad or pedal will make the port
active, for example it will provide input to the assigned
handpiece.
OFF: pressure on the footswitch pad or pedal will make the
port inactive, for example it will not provide input to the
assigned handpiece.
HIGH: pressure on the footswitch pad will cause the handpiece
to operate in the high-speed range or mode.
LOW: pressure on the footswitch pad will cause the handpiece
to operate in the low speed range or mode.
SPEED INCREMENT: pressure on the footswitch pad will
increment the set point speed.
SPEED DECREMENT: pressure on the footswitch pad will
decrement the set point speed.
IRRIGATION INCREMENT: pressure on the footswitch will
increment the pump flow set point and increase irrigation to the
handpiece cutting accessory.
IRRIGATION DECREMENT: pressure on the footswitch will
decrement the pump flow set point and decrease irrigation to
the handpiece cutting accessory.
IRRIGATION ON/OFF: pressure on the footswitch will toggle
the irrigation pump on and off. If pressed and held, the
irrigation pump can be turned on without running the
handpiece. Once started, the irrigation pump can be stopped
by pressing the pad again or by starting the handpiece.
PUMP FLUSH: pressure on the footswitch will turn the
irrigation pump on at the flush rate. Once started the irrigation
pump may be stopped by pressing the pad again or by starting
the handpiece.
CHANGE PORT: pressure on the footswitch pad will change
the footswitch assignment to another active handpiece.
ONE TOUCH: acting like an on/off toggle, tap the footswitch
pad to operate the handpiece at the maximum speed. Tap the
footswitch pad again to stop the operation.
VARIABLE: varying pressure on the footswitch pad will cause
the handpiece speed to vary.
FIXED: pressure on the footswitch pad will cause the
handpiece to operate at a constant speed as set on the HOME
screen.
Touch the arrow buttons to scroll
through the footswitch options.
The available options are
handpiece specific. Touch the OK
button to select the option.