User's Manual

21
DRAFT
To Adjust the Console Settings: - Standard and Advance User Mode
From the Adjust Options screen, touch the Adjust Console Options button to adjust the volume, brightness and contrast.
Adjust Volume
Adjust the percent of
console volume [0 to 100%]
by touching the slide bar or
the plus, minus buttons
Adjust Contrast
Adjust the percent of
console display contrast [0
to 100%] using the slide bar
or the plus, minus buttons
Adjust Brightness
Adjust the percent of
console display brightness
[0 to 100%] using the slide
bar or the plus, minus
buttons
Adjust Audio/Video Controls - Standard Mode
To Adjust Irrigation Pump Settings:
NOTE: If using the irrigation pump, assemble the irrigation pole to the mounting bracket located on the back of the
console. Hang the irrigation bag from the pole. Install the irrigation cassette into the console pump. Attach the
irrigation clips onto the handpiece and connect the tubing.
From the HOME screen, touch the Adjust Irrigation Pump Settings button to apply an irrigation flow rate to a handpiece
cutting accessory. AUTO PRIME and FLUSH options are also provided.
Adjust Irrigation
% Setting
Touch the slide
bar or plus,
minus buttons to
adjust the flow
rate percentage
of irrigation [1 to
100% or OFF] to
the cutting
accessory of the
handpiece.
At 100%,
irrigation to the
handpiece is set
to maximum
flow rate. The rate
decreases as the
setting approaches
zero.
FLUSH Tubing Set - Touch and
hold to provide a continuous flow
rate of 300 ml/min with no irrigation
clip and a minimum pressure of 30
psi when the tubing is blocked.
AUTO PRIME
Tubing Set
Touch to activate
the priming
sequence of the
installed irrigation
cassette.
The tubing will be
primed from the
pump head to the
irrigation tip clip
within 20 seconds.
The button will
remain depressed
during the timed
operation.
Select Handpiece
Touch one or all of the handpiece buttons to allow activation of the irrigation
pump. If no buttons are selected, the pump will not operate.
Activation will apply an irrigation flow rate to the handpiece cutting accessory.
The flow rate will depend on the type of handpiece.
Activating a handpiece will gain control of the console irrigation. While the first
activated handpiece is running, activating a second or third handpiece will not
change pump flow, but may reduce irrigation to the first handpiece if a Y-splitter
is used. To gain control of console irrigation using a second or third handpiece,
activate the second or third handpiece while the first handpiece is not running.