Problem
Sporadic electrical interference is
experienced.
Action
Turn off all electrical equipment not in
use in the operating room. Relocate
electrical equipment; increase spatial
distance. Plug the console and other
operating room equipment into
different outlets.
Troubleshooting Guidelines*
*The CORE Console is not field repairable. In case of operating difficulties, return this product to Stryker Instruments for repair. For
more information, contact your Stryker sales representative or Customer Service at 1-800-253-3210.
Interval
Prior to Use.
Activity
Inspect console for damage or signs of wear. DO NOT use if damage is apparent.
Periodic Maintenance
Storage and Handling
To ensure the longevity, performance, and safety of this equipment, package in original package materials when storing
or transporting.
Cleaning and Sterilization Recommendations
WARNINGS:
Clean and sterilize handpieces before first and every use.
Cutting accessories are single use only. Sterile only if package is unopened and undamaged. DO NOT resterilize or reuse.
CORE Console, Power Cord, Cable(s) and Footswitch Cleaning Instructions
CAUTIONS:
DO NOT sterilize the CORE Console or power cord.
DO NOT allow moisture in any CORE Console opening, or power cord, handpiece cable and footswitch electrical connection.
DO NOT use an aerosol spray directly on the CORE Console display.
1. Lightly wipe the CORE Console, cord, handpiece cable and footswitch surfaces with a soft cloth dampened with a non-abrasive,
hospital disinfectant or mild detergent and water.
2. Apply glass cleaner to a soft cloth and wipe the CORE Console display.