17.5 Appendix E: Power Cords
17.5.1 Power Cord General Specifications
Current Rating: 10 A
Voltage Rating: 250 VAC minimum
Frequency Rating: 50/60 Hz
Copper Conductor Size
Rating:
3 X 1.00 mm
2
≤ Conductor
Size < 3 X 1.50 mm
2
Connector Type: IEC 60320 C13
Cord Lengths:
3.0 m, 2.5 m
NOTE: The 2.5 m cord is not
for use in Canada or the US.
Temperature Rating: 0 °C to 70 °C minimum
Flammability Rating:
UL 94 V-2 minimum, IEC
60332-1
Cord Type:
SJT, H05VV-F, HVCTF, RVV
or equivalent (unshielded)
Dielectric Withstand:
1500 VAC for 60-seconds
between Line/Neutral and
Protective Earth
Mains:
Plug shall have a ground/
earthing pin
Certification:
All applicable in-country
medical electrical
requirements
17.5.2 Additional Power Cord Requirements
The Canadian and US power supply cord shall have
a tag or label in English and French indicating that
“GROUNDING RELIABILITY CAN ONLY BE ACHIEVED
WHEN EQUIPMENT IS CONNECTED TO AN EQUIVALENT
RECEPTACLE MARKED ‘HOSPITAL ONLY’ OR ‘HOSPITAL
GRADE’ “ or equivalent wording.
Agency Approval: CSA Certified for Canada and US
or UL Recognized for Canada and US.