User's Manual

24 www.stryker.com
EN 5400-052-700-EN Rev-D
17.8 Appendix H: Errors and Notifications
NOTE: Items within [brackets] in the following table are variables. For assistance, contact Stryker.
CODE TITLE CORRECTIVE ACTION
00000005 Motor Actuation
Error
The console only allows two motors to operate simultaneously.
00000006 Motor Actuation
Error
Connected motors cannot operate simultaneously.
00000007 Motor Actuation
Error
[Motor] has multiple active inputs. Only the first input detected is used. If the motor is
inactive, release all inputs and try again.
00001000
00001001
Motor
Functional Error
[Motor] or cord connected to Port [port] requires service. Report error code to Stryker. See
the Contact Information section of the Instructions For Use provided with the equipment to
contact Stryker.
00001002 Console
Functional Error
Report error code to Stryker. See the Contact Information section of the Instructions For Use
provided with the equipment to contact Stryker.
00001003 Motor Actuation
Error
Reset [motor] actuation. Release the motor button, trigger, or handswitch lever to reset. If
the error persists, report error code to Stryker. See the Contact Information section of the
Instructions For Use provided with the equipment to contact Stryker.
00001004 Motor
Compatibility
Error
Motor connected to Port [port] is not supported. See the For Use With section of the
Instructions For Use provided with the equipment.
00001005 Motor/Console
Functional Error
Error detected with [motor] connected on Port [port]. Remove [motor] from Port [port]; use a
different motor. If the error persists, report error code to Stryker. See the Contact Information
section of the Instructions For Use provided with the equipment to contact Stryker.
00001007 Motor
Functional Error
Error detected with [motor] connected to Port [port]. Remove [motor] from Port [port]. If the
error is resolved, return [motor] and cord to Stryker. If the error persists, report error code
to Stryker. See the Contact Information section of the Instructions For Use provided with the
equipment to contact Stryker.
00001008 Motor
Temperature
Error
[Motor] connected to Port [port] has reached an elevated operating temperature. Allow
handpiece to cool before restarting.
00001009 Motor
Temperature
Error
[Motor] or cord connected to Port [port] requires service. Report error code to Stryker. See
the Contact Information section of the Instructions For Use provided with the equipment to
contact Stryker.
0000100A Motor Service
Requirement
[Motor] requires service. Return handpiece to Stryker for service. See the Contact Information
section of the Instructions For Use provided with the equipment to contact Stryker.
0000100B Motor Stall
Error
Excessive load detected on Port [port]. Report error code to Stryker. See the Contact
Information section of the Instructions For Use provided with the equipment to contact Stryker.
0000100E Cutter Speed
Notification
Ensure [motor] speed does not exceed specified attachment limitations. Failure to do so may
result in user and/or patient injury.
00002000
00002001
Footswitch
Functional Error
Footswitch connected to port [port] requires service. Report error code to Stryker. See the
Contact Information section of the Instructions For Use provided with the equipment to contact
Stryker.
00002002 Console
Functional Error
Report error code to Stryker. See the Contact Information section of the Instructions For Use
provided with the equipment to contact Stryker.
00002003 Footswitch
Actuation Error
Reset Footswitch [port] actuation. Release the footswitch button or pedal to reset the
actuation. If the error persists, report error code to Stryker. See the Contact Information
section of the Instructions For Use provided with the equipment to contact Stryker.
0000221359, Rev. D Effective Date: Jun 1, 2017 10:00:55 AM
Print Date: Jun 08, 2017 08:54:46 AM