User's Manual

5400-052-700-EN Rev-D EN
www.stryker.com 25
CODE TITLE CORRECTIVE ACTION
00002004 Footswitch
Compatibility
Error
Footswitch connected to Port [port] is not supported. See the For Use With section of the
Instructions For Use provided with the equipment.
00002005 Footswitch/
Console
Functional Error
Error detected with footswitch on Port [port]. Remove the footswitch from Port [port]; use
a different footswitch. If the error persists, report error code to Stryker. See the Contact
Information section of the Instructions For Use provided with the equipment to contact Stryker.
00002006 Footswitch
Functional Error
Footswitch connected to port [port] requires service. Report error code to Stryker. See the
Contact Information section of the Instructions For Use provided with the equipment to contact
Stryker.
00003001 Console
Functional Error
Reset console. If the error persists, report error code to Stryker. See the Contact Information
section of the Instructions For Use provided with the equipment to contact Stryker.
00003002 Console
Functional Error
Irrigation pump is nonfunctional. Reset console. If the error persists, report error code to
Stryker. See the Contact Information section of the Instructions For Use provided with the
equipment to contact Stryker.
00003003 Console
Functional Error
RFID is nonfunctional. Reset console. If the error persists, report error code to Stryker. See
the Contact Information section of the Instructions For Use provided with the equipment to
contact Stryker.
00003004
00003005
Console
Functional Error
Reset console. If the error persists, report error code to Stryker. See the Contact Information
section of the Instructions For Use provided with the equipment to contact Stryker.
00003006 Console
Functional Error
Port illumination error occurred. Continue without illumination or report error code to Stryker.
See the Contact Information section of the Instructions For Use provided with the equipment
to contact Stryker.
00003007
00003008
00003009
0000300A
0000300B
Console
Functional Error
Reset console. If the error persists, report error code to Stryker. See the Contact Information
section of the Instructions For Use provided with the equipment to contact Stryker.
0000300C Motor/Console
Functional Error
Error detected with [motor] on Port [port]. Remove [motor] and cord from Port [port]; use a
different motor or cord. If the error persists, report error code to Stryker. See the Contact
Information section of the Instructions For Use provided with the equipment to contact Stryker.
0000300D Console
Functional Error
Report error code to Stryker. See the Contact Information section of the Instructions For Use
provided with the equipment to contact Stryker.
0000300E Console
Functional Error
Ensure irrigation cassette is installed properly.
00003010 Motor
Attachment
Error
Use a Stryker approved cutter. If the error persists, report error code to Stryker. See the
Contact Information section of the Instructions For Use provided with the equipment to contact
Stryker.
00003011 Irrigation
Cassette Error
Use a Stryker approved irrigation cassette. If the error persists, report error code to Stryker.
See the Contact Information section of the Instructions For Use provided with the equipment
to contact Stryker.
00003012
00003013
Console
Functional Error
Report error code to Stryker. See the Contact Information section of the Instructions For Use
provided with the equipment to contact Stryker.
00003014 Console
Functional Error
Error with audio function. Reset console. If the error persists, report error code to Stryker. See
the Contact Information section of the Instructions For Use provided with the equipment to
contact Stryker.
00003015 Console
Functional Error
Power supply over temperature. This may result from excessive use of motor duty cycle.
Report error code to Stryker. See the Contact Information section of the Instructions For Use
provided with the equipment to contact Stryker.
0000221359, Rev. D Effective Date: Jun 1, 2017 10:00:55 AM
Print Date: Jun 08, 2017 08:54:46 AM