17.10 Appendix J: Electromagnetic Compatibility
Guidance and manufacturer's declaration - electromagnetic emissions
The CORE 2 Console is intended for use in the electromagnetic environment specified below. The customer or the user of the CORE 2 Console should
assure that the console is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The CORE 2 Console uses RF energy only for the internal function. Therefore, the RF emissions
are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The CORE 2 Console is suitable for use in all establishments other than domestic, and may be
used in domestic establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes, provided the following warning
is heeded:
Harmonic emissions
IEC 61000-3-2
Class A
WARNING: This equipment/system is intended for use by healthcare professionals only.
This equipment/system may cause radio interference or may disrupt the operation of
nearby equipment. Mitigation measures may be necessary, such as reorienting or relocating
the CORE 2 Console or shielding the location.
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer's declaration - electromagnetic immunity
The CORE 2 Console is intended for use in the electromagnetic environment specified below. The customer or the user of the CORE 2 Console should
assure that the console is used in such an environment.
NOTE: The values provided in the table below have changed due to 60601-1-2 4th Edition requirements.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±2, ±4, ±6, ±8 kV Contact
±2, ±4, ±8, ±15 kV Air
±2, ±4, ±6, ±8 kV Contact
±2, ±4, ±8, ±15 kV Air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic material,
the relative humidity should be at least 30%.
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV at 100 kHz repetition
frequency for power supply lines
±1 kV at 100 kHz repetition
frequency for input/output lines
±2 kV at 100 kHz repetition
frequency for power supply lines
±1 kV at 100 kHz repetition
frequency for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±0.5, ±1 kV line(s) to line(s)
±0.5, ±1, ±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
0% U
T
(100% dip in U
T
)
for 0.5 cycle at 0°, 45°, 90°, 135°,
180°, 225°, 270°, and 315°
0% U
T
(100% dip in U
T
)
for 1 cycle at 0°
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 & 30 cycles at 0°
<5% U
T
(>95% dip in U
T
)
for 5 s
0% U
T
(100% dip in U
T
)
for 5 s
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
0% U
T
(100% dip in U
T
)
for 0.5 cycle at 0°, 45°, 90°, 135°,
180°, 225°, 270°, and 315°
0% U
T
(100% dip in U
T
)
for 1 cycle at 0°
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 & 30 cycles at 0°
<5% U
T
(>95% dip in U
T
)
for 5 s
0% U
T
(100% dip in U
T
)
for 5 s
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the CORE 2 Console requires continued
operation during power mains interruptions, it
is recommended that the CORE 2 Console be
powered from an uninterruptible power supply
or a battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m, 30 A/m at 50 and 60 Hz 3 A/m, 30 A/m at 50 and 60 Hz Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.