User's Manual

46 6390-009-001 REV B www.stryker.com
Return To Table of Contents
Powered Operations Instructions
Between
positions or
in sleep mode
Cot is locked
in position
Ready to
load cot
z
z
z
LE D B ehavior
Load Unload
1
2
4
1
2
3
4
POWERED OPERATIONS
System on
Battery is
charging
System Error
Use Manual
Operations
Low battery
3
CAPACITY
870
1 of 2
Locate this Powered Operation instruction label (6390-001-498) on the inside door panel or wall of the vehicle patient
compartment where visible to users.
To unload Power-LOAD:
1. Press and hold the release lever at the foot end of the Power-LOAD system and pull to remove the cot from the
vehicle patient compartment.
2. Press and hold the extend (+) button on the cot control switch to extend the cot undercarriage to the set load height
with the cot wheels on the ground.
3. Press the red release button on the cot control switch to unlock the cot from Power-LOAD.
4. Raise the lifting arms and guide the trolley into the vehicle patient compartment until the arms are far enough in
to not interfere with the vehicle doors.
5. For additional unloading instructions, see “Unloading a Power-PRO Cot from a Vehicle (Model 6500/6506 &
6510/6516 with the Power-LOAD Option)” on
page 57.
To load Power-LOAD:
1. Raise the lifting arms to guide and pull the trolley out of the vehicle patient compartment.
2. Push the cot into Power-LOAD until the cot load wheel pins lock into position. Ensure that the cot is aligned with
the lifting arms when loading.
3. Press and hold the retract () button on the cot control switch to fully retract the cot undercarriage until the cot
is supported.
4. Push the cot into the vehicle patient compartment until the lifting arms lower and the cot locks into Power-LOAD.
5. Ensure that the cot is locked into Power-LOAD by firmly pulling on the foot end of the cot.
6. For additional unloading instructions, see “Loading a Power-PRO Cot into a Vehicle (Model 6500/6506 & 6510/6516
with the Power-LOAD Option)” on
page 55.
See “User Controls and LED Indicators” on
page 41 for button and LED locations and descriptions.
Note: If Power-LOAD will not be in use for a week or more, press the main power button to turn the unit off and avoid
draining the battery.