DATA LOGGER Felix Storch, Inc. │770 Garrison Ave. Bronx, NY 10474│ Tel: (718) 893 3900│ Fax: (844) 478 8799 USER MANUAL (DL2) calibration@summitappliance.
FEATURES Data Logger displays current temperature. The unit will provide a visual and audio alert when temperature rises above or falls below the high and low set points. The min/max feature is designed to monitor and store the highest and lowest readings until the memory is cleared, or removal of battery. The temperature sensor is enclosed in a glycol-filled bottle, protecting it from rapid temperature changes when refrigerator/freezer door is opened.
Package Contents Data logger Temperature sensor in a glycol-filled bottle Instructions manual Magnetic sticker (w/instructions) Rechargeable Lithium battery 4 GB Memory stick Power adapter NIST-traceable calibration certificate Parts and Controls/Features °C/°F Press the °C/°F button to change the temperature display scale from Fahrenheit (oF) to Celsius (oC) and back. Up/Down keys to change settings.
Power Adapter: Connect the unit to an AC power outlet. Data Logger rear view Do not use force to connect the probe or the power adapter plugs. The power adapter plug is different from the probe plug. Power adapter port Temperature probe port LCD Display Description 1 Low battery symbol.
Set Up Battery: To replace the battery, unscrew the battery compartment cover located on the top of the unit and install the battery. Follow polarity (+/-) diagram below. Replace the battery cover. The unit will beep and all segments of the LCD will be activated.
Operation Programming the Data Logger Year From the main display screen: Hold the MODE key for 3 seconds, then press the Mode key TWICE to enter the Year setting mode. Use the up and down arrows to adjust the year accordingly. Press MODE key ONCE to confirm the settings. Month From the main display screen: Hold the MODE key for 3 seconds, then press the Mode key 3x to enter the Month setting mode. Use the up and down arrows to adjust the month accordingly. Press MODE key ONCE to confirm the settings.
HIGH AND LOW ALARM By default, the high and low alarm settings are 8 oC and 2 oC respectively. To reset high alarm and low alarm temperature settings, follow instructions below. This is the high temperature limit. The current temperature is considered out-of-range if High Alarm Temperature the probe reads a temperature equal to or above the high temperature limit. Setting From the main display screen: Hold the MODE key for 3 seconds to enter High Alarm temperature setting mode.
To Use Once the Data Logger has been programmed, place the temperature sensor (in glycol bottle) in the location to be monitored, such as inside a refrigerator or freezer. Data Logger may be placed on top of unit with LCD display easily visible and the alarm audible. Data Logger displays internal temperature of unit being monitored, as well as maximum and minimum temperatures reached.
CSV File To download data, USB drive is ejected safely and connected to a computer. Open file(s) in Microsoft Excel or any .CSV compatible program.
Troubleshooting Displays ‘NP’ The temperature sensor is not installed correctly. Display screen not working Make sure the AC adapter and Li-ion battery are installed correctly. “Low battery” indicator flashing Battery may need to be recharged. Press the Rec/Stop key and make sure the L symbol appears on the left side of the display. The logger will stop logging if AC power is removed and rechargeable battery is not connected or not charged.
Recalibration It is recommended that Felix Storch, Inc., temperature monitoring products be calibrated every two years to ensure proper function and accurate measurements; however, your quality system or regulatory body may require more frequent calibrations. To schedule your recalibration, please contact our ISO/IEC 17025:2005 calibration laboratory accredited by Perry Johnson Laboratory Accreditation, Inc., at calibration@summitappliance.
AccuCold, Div of Felix Storch, Inc. • 770 Garrison Ave.