Owner's Manual
Page 12
below the LO set point, the alarm will sound for 1 minute and issue a
3-second repeater beep every minute thereafter, for up to 12 hours.
The alarm will continue to sound even if the current temperature
display returns to an “in range” condition.
To temporarily silence the alarm, press either the HI or LO button on
the back of the box. The box is still active and the alarm will sound
again if the temperature goes out of range. To permanently silence the
alarm, slide the ALARM ON/OFF switch to the OFF position.
RESETTING THE HIGH/LOW ALARM
Use a pointed object to push the RESET button on the back of the unit.
It is necessary to reset the unit whenever a change is made to °C/°F
temperature, FAST/NORMAL sampling rate, or if the probe sensor is
plugged in or removed.
NOTE: HI/LO temperature alarm settings and MIN/MAX memories will
be cleared after resetting the unit.
RESETTING MIN/MAX MEMORIES
There are two ways to reset the memories:
1. To reset both Minimum and Maximum memories, press the
MEMORY CLEAR button.
2. To reset individual Minimum and Maximum memories, press either
the MIN or MAX button (on back of unit) once to reset the respective
memory. The other memory will remain unaffected.
NOTE: You must be in NORMAL DISPLAY MODE to reset the memories.
DISPLAYING °F OR °C
To display the temperature readings in Fahrenheit (°F) or Celsius (°C),
slide
the switch on the back of the unit to °F or °C.
NOTE: The unit must be reset after changing the temperature scale.
SELECTING THE SAMPLING RATE
To select normal sampling rate (every 60 seconds) or fast sampling
rate (every 10 seconds), slide the switch on back of unit to NORMAL or
FAST.
NOTE: The unit must be reset after changing the sampling rate. Using
the NORMAL sampling rate will extend battery life.
If your model has a front lock, the lock has a feature called Automatic
ALARM INSTRUCTIONS (cont’d)
KEY EJECT LOCK (Selected models)
Felix Storch, Inc. www.summitmedicalrefrigerators.com
Page 9
Unless you specify otherwise, your appliance has a
mechanical (dial) thermostat. For FF and SCR models, the
thermostat is located inside the unit, on the right-hand side
wall, at the top. For model FS62LBI7MEDSC, the thermostat is
on the exterior front of the unit, on the right side of the
tabletop face. For model FS62LBI7MEDSCADA, it is on the rear
of the unit, near the top.
HOW TO ADJUST THE THERMOSTAT
The thermostat controls the temperature inside the refrigerator. Turn
the thermostat knob clockwise from the STOP (0) position toward the
higher numbers for colder
temperatures.
Temperature Controls
Higher settings on the knob (higher numbers) mean colder
temperatures in all compartments of the appliance. The temperature
in the interior of the appliance may even drop below 32°F. Use higher
thermostat settings only when required or recommended. When the
ambient temperature is normal, we recommend a medium setting
.
Changes in the ambient temperature affect the temperature in the
interior of the appliance. (Choose the setting of the thermostat knob
with care.)
In STOP (0) position, the appliance does not operate (the cooling
system is switched off), yet the power supply is not interrupted (the
light is on when you open the refrigerator door).
The temperature in the refrigerator and in the freezer compartment
also depend on how often you open the door.
NOTE: In all small refrigerators and freezers without interior fans,
temperatures are coldest near the bottom, and occasionally near the
rear. Place contents accordingly. It is best not to operate these units
mostly empty. If temperatures are erratic, try putting some water
TEMPERATURE SELECTION
www.summitmedicalrefrigerators.com Felix Storch, Inc.