generated by the laser equipment being used. Solvents and flammable
solutions used for cleaning and disinfection should be volatilized before
using the equipment. There is a risk the internal gas may be ignited.
WARNING: Diode lasers are attracted to melanin and hemoglobin.
Power levels are inversely related to the amount of pigmentation and
vascular of the targeted tissue. As pigmentation varies from patient to
patient, start with the lowest power setting and increase as needed to
accomplish the procedure in tissues which have lower melanin and
vascular content; Use anesthetic as needed for patient comfort.
WARNING: To avoid risk of electrical shock, AC/DC adapter must
only be connected to a properly grounded power source!
2.5 Safety Classifications
The following safety classification applies to the P4H:
Laser radiation: Class 4 laser products
Type of Electric shock protection: Class 1
Degree of Electric shock protection: Type B
Indication of enclosure protection: IPX0
2.6 Security Issues
Inform Summus Medical Laser or the designated dealer directly about the
safety problems discovered during the use of this product.
2.7 Security Features
The P4H provides a range of safety monitoring and safety functions for
monitoring the operating status of the system and emergency interruption of laser
output. The purpose and operation of these safety devices should be familiarized
when using the diode laser equipment.
Power Switch
Used as a line switch for separating the device from the main power supply:
O = OFF, | = ON.
Login Password
The login interface is prompted after the equipment is powered on requesting