the user's 4-digit security PIN identification for permissions.
Emergency Stop
The emergency stop switch is used to immediately stop the laser emitting
output. When the device is powered on and operating, if an abnormal function
(such as no display, and light abnormality, etc.) occurs during the course of
treatment, immediately press the emergency stop switch to stop the laser output.
Identify and correct the cause prior to resuming use. The user can reset the
EMERGENCY STOP switch by pulling the switch out and away from the
equipment.
Remote Interlock
If the interlock is not inserted into the connector on the device, all electrical
power to the controls and laser components is terminated. The safety interlock
MUST be inserted before the device can power on. Interlock should be installed at
each designated therapy room entrance and connected to the remote interlocking
connector on the rear panel of the device.
The remote interlock should normally be connected to an interlock (closed
switch) on the entrance door of the therapy room. The remote-control interlock
head should be connected to the remote-control interlock interface. When the
therapy room door is opened, the equipment can no longer emit. An auxiliary
connector should be used to access the equipment if the room does not have an
interlock.
System Monitoring
The P4H is equipped with an emergency stop switch, remote interlock,
foot/finger switch, fiber, laser temperature and other system status functions. If the
system detects any associated anomalies or errors, the laser emitting output will
be stopped and an audible, visual alarm will alert the user device to an error.
Start Treatment (Standby/Ready) Button
The power or program selection settings can only be made in the standby
mode. When the system is in all normal conditions, press the START
TREATMENT (Standby/Ready) button to switch the device from Standby to
Ready mode. Activation of the finger switch will allow the equipment to enter the
laser emission state. Users are informed the equipment is emitting laser through a
beeping sound and a yellow light.
Indicator Lights