Inflammation
>5,000 Hz
Average power remains constant during emission in CW or Pulse respectively. Refer to
Recommended Treatment Protocol tables for light or dark skin below for average power.
Recommended Treatment Protocol for Light Skin
Fitzpatrick scale 0 through 3
Intended
Body
part
Average
Power,
Watts
Power
Density,
Watts/cm2
Treatment
time,
seconds
Total
Dose
delivered,
J
Treatment
Area size,
cm2
Dosage,
J/cm2
Wavelength
Cervical
7
0.36
360
2520
400
6.3
*
Thoracic
8.5
0.43
420
3570
500
7.14
*
Lumbar
12
0.61
420
5040
500
10.08
*
Shoulder
8
0.41
360
2880
400
7.2
*
Elbow
6
0.31
180
1080
200
5.4
*
Forearm
5
0.25
240
1200
300
4
*
Wrist
4
0.20
120
480
100
4.8
*
Hand
4
0.20
90
360
100
3.6
*
Finger
2
0.10
45
90
25
3.6
*
Hip
10
0.51
420
4200
500
8.4
*
Thigh
9
0.46
420
3780
450
8.4
*
Knee
8
0.41
360
2880
400
7.2
*
Leg
7
0.36
300
2100
300
7
*
Ankle
6
0.31
240
1440
200
7.2
*
Foot
5
0.25
150
750
100
7.5
*
Toe
2
0.10
45
90
25
3.6
*
* Depending on laser device model (P1 = 650, 810, 980nm or P4 = 650, 810, 915, 980nm)