8.3 Alarm Function
The Platinum Series P4H
Laser monitors the operating status of the system
in real time. Sounds, indicators, and graphical warning signs alert the system to
unusual events with different priorities while interrupting the laser outputting.
Alarm Types
Triggering conditions
Alarm release method
High
Temperature
Alarm
1.Laser temperature exceeds
45 ° C
2. Fiber coupler temperature
exceeds 60 ;
3. Temperature sensor failure
1. If the alarm is caused by high
temperature, power off the
device. Air cooling device for
~10 minutes will return to
normal after the reboot. Do not
cover the opening of the air flux.
2. Please contact the company
after-sales maintenance.
Remote
Interlock
Alarm
The remote interlock is not
inserted correctly or is not
plugged into the device.
Check the rear panel of the
device and whether the remote
interlock is correctly inserted
into the device. If the remote
interlock is inserted into the
device and the alarm condition
is still unresolved, please
contact the company for
after-sales maintenance.
Hand Piece
Switch Alarm
The hand piece switch is not
properly inserted or is not
inserted into the device.
1. Check the front panel and
whether the hand piece switch
is correctly inserted into the
device. 2. If the handle switch is
properly inserted into the
device, and the alarm status is
still unresolved, please check
the handle switch button is
pressed or not.
3. Check the above operation, if
there is still alarm status, please
contact the company after-sales
maintenance.
Laser Alarm
Safety interlock failure
1. Please contact the company
for after-sales maintenance.
Fiber Alarm
Fiber connector is not inserted
into the laser window.
Micro - switch of laser window
fails.
Insert the fiber connector into
the laser window of the device.
If there is still an alarm
condition, please contact the
company after-sales
maintenance.