9 POST TREATMENT CLEANING
Appropriate steps shall be taken by the operator between treatments of the laser
device and components to ensure the distal parts coming into direct or indirect
contact with the patient such as the applicator treating head/ hand piece tips
receive a thorough cleaning process.
Thorough manual cleaning shall be completed by the operator for all reusable
components to assure all surfaces are completely free of any organic or inorganic
material to prevent transmission of potential bioburden.
Wear appropriate Personal Protective Equipment (PPE) when performing
cleaning and disinfecting procedures.
WARNING: Following treatment, switch off the device and disconnect the
power cable from the power supply.
WARNING: Unit and accessories are not sterilizable.
9.1 Cleaning of the Distal Hand Piece Tip
WARNING: The applicator treating head/ hand piece tips may be warm
immediately after treatment. Allow to air cool.
Do not twist the hand piece for possible damage to the optic fiber.
Remove the zoom distal part from the hand piece by gently pulling the tip from
the hand piece for cleaning.
Make sure that no dust or dirt enter the optical fiber socket or the hand piece
optical system as these may permanently damage the unit.