6) Please contact the company or the designated dealer to repair and
maintain the device when the power is reduced. Do not disassemble the
device without guidance of Summus Medical Laser personnel or
designated dealers to avoid injury to the patient and damage to the device
caused by possible harmful laser radiation.
7) Summus Medical Laser recommends carrying out routine inspection and
maintenance and power calibration every year under the guidance and
operation of authorized company personnel or designated dealers to
avoid possible injury or damage to the device from harmful laser radiation.
10.2 Routine Inspection
Clinicians can routinely check the device under the guidance of the dealer or
on their own to ensure that the device is working properly. The general check
contents are as follows:
1) Whether the safety device is normal: Safety interlock, finger switch, optical
switch, emergency stop switch
2) Whether the finger switch signal is normal: Emit the laser in the READY
state
3) Whether the sound or Indicator light is normal
4) Whether the operation of touch screen is normal
5) Whether the label is affixed firmly
6) Whether the laser power is within the normal range
7) Whether the optical fiber is normal (i.e. bent, broken) before each
treatment)
If national or local legal regulations require additional safety checks for
the laser unit, these regulations must be complied with and the corresponding
checks shall be performed by the user.
10.3 Laser Power Calibration
Warning: When the power output has changed and the speed exceeds ± 10%
of the set power, please contact a Summus Medical Laser designated
representative for required power calibration. Calibrations must be under the
direction of our companies authorized personnel to avoid possible injury or