Product Overview
LCV: AC Current, AC Voltage Logger
The LCV data logger is a precision instrument for recording AC current and AC voltage.
AC current is sampled using the supplied clamp probe. The logger can store up to 21,500
data points, when it is setup to record either AC current or AC voltage (but not both). The
logger can store up to 10,750 points, when it is setup to record both AC current and AC
voltage. Each logger is factory calibrated. This means that the clamp probes are not
interchangeable.
LCV
Current & Voltage Data Logger
SEALED UNIT PARTS CO., INC.
PO Box 21, 2230 Landmark Place, Allenwood, NJ 08720 USA • Phone: 732-223-6644 • Fax: 732-223-1617 • www.supco.com • info@supco.com
LOGiTLOGiTLOGiT
Sample Point Capacity: 10,750 points for Current & Voltage
21,500 points for Current OR Voltage
Sample Rate: 1 second to 18 hour
Alarms: Over and under alarms for current and voltage Visual alarm
indication
Current Range: 0 to 300 Amps AC
Current Accuracy: +/- 5%
Current Resolution: 0.1 Amp AC
Voltage Range: 0 to 500 Volts AC
Voltage Accuracy: +/- 1 V AC
Voltage Resolution: 0.1 V AC
Time Accuracy: +/-100 ppm @75°F
Operating Range:
Temperature: 32°F to 140°F (0°C to 60°C) using Alkaline Battery.
-40°F to 150°F (-40°C to 65°C) using Lithium Battery.
Humidity: 0% to 95%, Non-condensing
Storage Temperature: -40°F to 170°F (-40°C to 77°C)
Calibration: Single point offset calibration available through software.
Current and Voltage are factory calibrated.
Dimensions: 3” x 2.5” x 1.1” (7.6 cm x 6.4 cm x 2.8 cm)
Weight: 2.5 Oz (71g)
Power Source: 9 V Alkaline or Lithium battery.
1 year warranty
PC SOFTWARE OPERATES WITH ALL LOGIT LOGGERS (P/N LLS NOT INCLUDED)
SPECIFICATIONS
LOGiT Common Specifications
Sampling Interval User selectable from 1 second to 18 hours, in 1 second
increments.
Interface COM (RS232) serial port. Software supports COM1
through COM8. Optional USB interface.
Alarms Visual alarms with user selectable low and high thresholds
for every channel.
Alarm Memory Up to 50 alarm events. Alarm start time and alarm
duration is stored.
Data Security Locking sliding door prevents physical access to the logger
data. Unique factory set logger id. Data collection session
counter.
Compatibility Windows 95 and above.
Minimum System
Requirements 8MB of RAM, 2M of disk space, 800 x 600 resolution, 1
free COM port or a free USB port (optional USB interface is
required).
Recording Start On door close, immediate after setup, or at a user specified
date and time.
Recording Mode Stop on memory full, or continuous recording with
memory rollover.
Real Time Status Visual indication through the status LED. User selectable
status display frequency and LED brightness. Software
real time status.
User Calibration Password protected calibration for each channel.
User Message Up to 30 characters.
Logger Channels User selectable. Disabling channels increases the storage
capacity of the other channels
Units US and Metric. User selectable units.
Data Format Custom format exportable to text or Excel spreadsheet.
Real Time Monitoring Logger can be permanently connected to the computer and
used as a real time, multichannel sensor with logging
capability. Battery does not drain when logger is
connected to the computer.
LCV Specifications
Sample Point Capacity 10,750 data points when AC current and AC voltage are
recorded together. 21,500 points when recording only AC
current or only AC voltage.
Alarms Over and under alarms for AC current and AC
voltage.
Calibration AC current and AC voltage are factory calibrated. User
single point offset calibration is available through software
for both AC current and AC voltage.
Operating Temperature 32°F to 140°F (0°C to 60°C) using Alkaline Battery.
-40°F to 150°F (-40°C to 65°C) using Lithium
Battery.
Time Accuracy +/-100ppm @75°F
Relative Humidity Range 0% to 95%RH non condensing
Current Range 0 to 300 Amp AC
Current Accuracy +/- 5%
Current Resolution 0.1 Amp AC
Voltage Range 0 to 500 Volts AC
Voltage Accuracy +/- 1 VAC
Voltage Resolution 0.1 VAC
Dimensions 3'' x 2.5" x 1.1'' (7.6cm x 6.4cm x 2.8cm)
Weight 2.5 Oz (71g)
Power Source 9V Alkaline or Lithium battery.
Battery Life (average use) 18 Months using Alkaline Battery.
36 Months using Lithium Battery.
Connecting the LCV to Electrical Equipment
Before connecting to any equipment follow these precautions:
• Make sure that the wires are securely attached to the connector and that no uninsulated
portion of the wire is visible after the connection is made.
• Make sure that there is no strain that could pull the leads out of the LCV connector.
•Make sure that the leads will not interfere with any moving pieces of equipment and will
not pull loose any wires or other connections.
To measure the voltage, connect the two voltage clip leads to the points at which the
voltage is being measured.
To measure the current, place the clamp probe around one of the conductors or wires in
which the current is being measured.
CAUTION!
Electrical connection to any type of operating equipment is Dangerous!
The LCV leads and current probe should only be connected when it has been determined
that the power is off and the equipment is safe to touch.
It is recommended that you always remove the LCV voltage leads before connecting
the LCV to your computer!
The LCV Current & Voltage Data Logger measures external
Single Phase Current and Voltage, simultaneously
(Use 3 LCV’s to measure 3 Phase Current and Voltage.)
•
•
One (1) Current Measurement
One (1) Voltage Measurement
Simultaneous Monitoring of Current and Voltage for:
Compressor and Fan Run Cycles
Troubleshooting Intermittent problems
Verification of Power Conditions
•
•
•
®