Operation Manual

7. Inlet and Outlet sections
7. Inlet and Outlet sections
The thermal measuring principle may be sensible to inlet and outlet
conditions. For this we recommend the following minimum straight inlet
and outlet sections to ensure an accurate measurement. Sensor should
be always installed upstream from obstacles like valves, filter,
reductions etc. In common the sensor should be installed as far as
possible away from any disturbances.
Note: If there is any combination of the below, the longest straight inlet
section must be maintained.
The S 421 sensor comes with its own measuring section and a straight
inlet section, nevertheless additional straight inlet sections must be
added to measuring sections to fulfill the minimum inlet requirements.
Please refer to the installation types below and select your additional inlet
section 'LA' from the table.
Note: The additional inlet sections depend on the pipe diameter. The units for
the length are in mm.
14 S 421