User's Manual

Document No.:
46 / 91
the scanning end interface (Figure 5-5-3 or Figure 5-6-3 or Figure 5-7-3), to enter the
system setting interface (Figure 5-12). Setting parameters include: Operation mode
selection, instrument calibration, interface language, automatic poweroff time, volume alert
preset, password management and system information. Click return key on the system
setting interface to return to the main interface. There are 7 options on the system setting
interface, and the functions are as follows:
6.10.1 Mode selection setting
Click the "User Mode" on the system setting interface to enter the operation mode
setting interface (Figure 5-12-1). There are three options: Expert mode, easy mode and
intelligence mode. It is expert mode by default. After the selection, the " " will
automatically appear in the setting item confirmation field behind the selected operation
mode. Click the return key , to return to the selected main interface.
When selecting the “expert mode”, during pre-scanning, the operator can observe the
bladder B-ultrasound image of the patient, move the probe to center the image and
maximize the section area of the bladder. During scanning, 12 real-time bladder
B-ultrasound images (Figure 5-5-2 and Figure 5-5-3) of the patient will be displayed on the
screen. After scanning, the volume value and the scanning projection of the bladder will be
displayed on the screen. At this time, click the scanning projection in the round frame to
view the 3D image of the bladder. Or click the 3D images of the bladder to view the
scanning projection.
When the projection is displayed in green, it means that the bladder is in the scanning area;
otherwise, it is in orange.
When selecting the “easy mode”, during pre-scanning, the operator can observe the
bladder section of the patient, move the probe to make the center of the bladder section to
the center line of the circle, and to maximize the bladder section area. During scanning, 12
bladder section diagrams (Figure 5-6-2 and Figure 5-6-3) will be displayed on the screen.
After scanning, the volume value and the scanning projection of the bladder will be
displayed on the screen. At this time, click the scanning projection in the round frame to