User's Manual

Document No.:
Attached page to the Specification
Instructions for Auto Calibrate:
The M series bladder scanner is patented in the fields of ultrasound imaging,
measurement algorithm and probe. Therefore, the M series bladder scanner has the
advantage of clinical application of no calibration required during the service life. For this,
we officially state that: In the lifelong clinical application of the M series bladder scanner and
the probe, as long as they are intact:
1. The operator does not need to calibrate the clinical measurement accuracy of the
bladder scanner before first measurement of the patient's bladder volume.
2. The operator does not need to calibrate the clinical measurement accuracy of the
bladder scanner during daily application.
If the hospital or any end user regards instrument calibration as fixed operation
procedure, the following steps may be used for calibration. Please note that: Model 616
ver.2014 Bladder Scanner Calibration Phantom for Caresono is the only applicable
calibration phantom model designated. Since the purchased phantoms provided by the
supplier are manufactured separately, the consistency cannot be guaranteed. The material
formulation of each phantom is slightly different, which will affect the measurement
accuracy of the phantom (will not affect the measurement accuracy of the bladder volume),
therefore, we conduct Auto Calibrate to Model 616 ver.2014 Bladder Scanner Calibration
Phantom for Caresono to ensure the measurement accuracy of the phantom.
The steps of Auto Calibrate are as follows:
Step I: Enter the last item "System Info" on the system information view interface (Figure 1),
and click the hardware version number (Hardware Version: M4-HD.1.0.0.X) of the display
instrument for 5 times continuously. Activate, and the Auto Calibrate switch light is on in
green. The calibration confirmation interface appears (Figure 2).
Step II: Operate according to the prompts on the calibration confirmation interface: 1.
Connect the wireless probe first. 2. Place the probe head into the phantom holder firmly. 3.
Do not remove the probe during the calibration. After the operation, press the Start key