User's Manual

Document No.:
5 / 90
image aligns with the green indication line in the middle of the screen.
Under pre-scan of the easy mode, observe the bladder sections and move the probe till the biggest
section is obtained. While achieving the biggest sectional view, make sure the centre of the section aligns
with the middle point of green indication line.
Under pre-scan of the intelligence mode, observe the position of the bladder projection. If the projection
image is off the centre of the crosshair, indication arrows ←↑→↓↗↘↙)will appear to guide the moving
direction of the probe. Move the probe by following the direction of the arrows till the projection image
reaches the centre of the crosshair. When the projection image is green, it means bladder centre has
been located. Otherwise it turns to orange.
Under all three operation modes, the brightness of the screen could be adjusted by putting the figure next
to the projection image and moving it upside down on the touch screen.
Remark:
During pre-scan some functions are locked including gender selection, patient information edition, storage
and print, as well as setting. Icons of those functions turn to grey. After the pre-scan is completed, those
functions get unlocked.
When the color of the icons is white, it means those functions are unlocked and can be used. When it
turns grey, the functions are locked and cannot be used.
6.4.2 Bladder Scan
After bladder is positioned, keep the probe still and press SCAN button. 3D probe starts to scan.
6.4.2.1 Expert Mode
During scan twelve B-Mode images are displayed on the right side of the screen. Serial number of the
images increases from 01 to 12. When it appears as 12/12, it means scan, image analysis and volume
calculation is completed.
Under gallery view, bladder volume and 3D bladder projection appears on the left side of the screen. the
twelve images are displayed on the bottom right of the screen side by side. Slide left or right to view all
twelves images. Tap on any of them, an enlarged image appears.
Under grid view, bladder volume and 3D bladder projection appears on the left side of the screen as same