Programming Instructions

Programming screen « USER » Page 58/78
General programming instructions
b) Calibration status
- Depending on the stability and measurement value of the sensor, the status screen may change as
below. It’s necessary that the measurement value is stable and in a “correct” range to validate the
calibration.
- Status 1
Very unstable sensor (Measurement value changes rapidly)
Sensor disconnected or Out of order (Measurement and Raw signal displays “- - -”)
- Status 2
Sensor being stabilized
- Status 3
Stabilized sensor
- Status 4
Authorized calibration
c) Offset calibration
Wait for the sensor stabilization
Enter the calibration value
Press “SAVE”
f) Slope calibration
Wait for the stabilization
Enter the calibration value
Press “Save”
g) Erase calibration
Press “Clear”
A confirmation request will be displayed.
There is no « analysis » of measurement value, erasure can be performed even if the sensor is unstable
or absent.
Status 1
Status 2
Status 3
Status 4
Sensor
Measure : 5.00 mg/l
Raw signal : 10.00 mA
Stability
Low Good
Offset
0 mA
Slope
100 %
Sensor
Measure : 5.00 mg/l
Raw signal : 10.00 mA
Stability
Low Good
Offset
0 mA
Slope
100 %
Sensor
Measure : 5.00 mg/l
Raw signal : 10.00 mA
Stability
Low Good
Offset
0 mA
Slope
100 %
Sensor
Measure : 5.00 mg/l
Raw signal : 10.00 mA
Stability
Low Good
Offset
0 mA
Slope
100 %