Installation Instructions

Input mode Page 45/67
Installation and starting instructions
« Disable the parameter » & « Activate the parametre » button
This button allows you to stop managing the parameters. Alarms, dosing, relays and
4…20mA outputs associated with this parameters will be stopped.
« History » button
This button displays the channel history. (See §VII.3 Graphical channel history display)
« Main sensor information » area
Type : Type of the main sensor
Scale : Scale of the sensor
Raw signal : Input measurement signal
« Dosing » area
List of dosing units assigned to the parameter :
FO1 : Output number
: Dosgin direction
PFM : Operating mode of the system
P.I.D : Dosage calculating mode
Setpoint button : Used to change the parameter setpoint
Time Max. button : Used to change the maximum dosing time
Disable the
parameter
Activate the
parameter
History
Type
:
Potentiometric
Scale :
0,00…14.00 pH
Raw signal :
Main sensor information
FO1 - - PFM - P.I.D
Dosing
Setpoint
Time Max.
Setpoint
7,20 pH
Min :0
Max
:14.00
Time Max.
OFF
Min : 0
Max :
1080
o Setpoint button
o
Time Max. button