Installation Instructions

Input mode Page 47/67
Installation and starting instructions
« Controlling » area in the case of an input Volume
List of controlling threshold
Dosing stop threshold button : Used to change the dosing stop
threshold.
o « Dosing stop threshold » button
« Calibration » area
List of the different calibration parameters associated with the
input.
Offset : Indicates the offset value has been set
Slope : Indicates the percentage of slope has been set
Calibration button : Used to change the calibration value.
o Offset (Example pH) :
Display the state of the sensor offset calibration
The more the bargraph becomes red, the more the calibration limit is reached.
This information can be a sign of sensor aging.
The offset is expressed in the unit of the physical sensor input. The min. and max.
values are different according to the sensors.
Calibration
Offset
0mV
Calibration
Slope
100%
Offset
-60mV
Dosing stop
OFF