Installation Instructions

Maintenance Page 19/24
Installation and commissioning instructions
Position the bead-filled cap under the probe and fill it back on the probe by turning clockwise
and locking the packing cap on the O-ring above the copper electrode.
3) Change of the copper counter-electrode
The change of the copper counter-electrode must be done every year. This period may vary depending
on water quality and flow rate on electrodes.
Please follow these steps to change the copper counter-electrodes:
Disconnect the power supply wires from the measurement loop.
Remove the probe from the measuring chamber (see above)
Take the probe vertically and undo the packing cap by turning clockwise, take care not to lose
glass beads.
Loosen the copper counter-electrode clockwise.
Collect and discard the seals. At each electrode change the seals must be changed. BE
CARREFUL NOT to damage the platinum or gold rod!
Replace new gasket and tighten new copper counter-electrode until O-ring is tight.
Reassemble the packaging cap with the glass beads.
After a change of counter-electrode the zero-point and slope calibration must be carried out as
described in § « VI Calibration ».
During all operations be careful not to touch and damage the gold or platinum rod.