User's Manual

Application Deployment 9 - 9
2. Click the LOAD button. The hex file(s) is downloaded to the device.
Figure 9-10
Download Complete Screen
3. On completion, press ENTER to return to the Bootloader main screen to select the next file to download.
4. To exit Bootloader, select Exit from the Bootloader main screen and press ENTER.
Bootloader Error Detection
While receiving data, Bootloader performs many checks on the data to ensure that the data is received
correctly. If an error is detected, Bootloader immediately aborts the download, and reports the error on an error
screen.
This error message screen displays until a key is pressed. Once the screen is acknowledged, Bootloader
returns to the main menu to wait for a new selection.
To find the probable cause of the error, use the error number and/or the error text displayed on the screen to
look up the error in Table 9-1.
Table 9-1
IPL Errors
Error Text
Error
Number
Probable Cause
Unknown error -1 A general error occurred. Retry the download. If the failure persists,
it is most likely due to a hardware failure; the MC18 requires
servicing.
Cancelled by user -2 The user canceled the download.
Can't open the source -7 An error occurred opening the source device (either radio card or
Serial port). Check source device connectivity and retry.
Can't open the
destination
-8 An error occurred opening the destination device (either flash ROM
or Power Micro). Retry the download. If the failure persists, it is most
likely due to a hardware failure; the MC18 requires servicing.
Can't read from the
source device
-9 The source device (either radio card or Serial port) could not be
read from. Check source device connectivity and retry.
Can't write to the
destination device
-10 The destination device (either flash ROM or Power Micro) could not
be written to. Retry the download. If the failure persists, it is most
likely due to a hardware failure; the MC18 requires servicing.