4 Mounting
•
Before installation remove the device from the packing and check for any possible
shipping damage!
•
Assemble the device on a clean and stable base. Do not distort during assembly! Use
the appropriate mounting devices for proper installation of the unit!
•
Use the templates printed on the device packing to mark the fastening bore holes.
•
The pressure measuring depends on position, therefore the mounting must be made
vertical and as possible on a vibration-free place (cable inlet and pressure connec-
tions down).
•
The pressure line´s connection should be with plastic-hose (in building), inside diam-
eter 4 / 5 mm. For a firm hold of the hose, its inside diameter must be 1 mm smaller
than the outside diameter of the hose nozzle (step spigot 5 / 6 mm).
•
Remove the connection cover for mounting, electrical connection and setting of the
measuring range. Close the lid again carefully before start-up (tightening torque of the
lid screws 1.1 Nm).
=5x(16)=5x(16)Vertical mounting
Gehäuse
housing
111 mm
Gehäuse / housing
106 mm
11 mm
83 mm
91 mm
Schraube
max 4,5 mm
screw max
4,5 mm
Bohrschablone
drilling template
Drilling template on packing
Operating Instructions MPG-..V Mounting
L-BAL-E248-GB 1515 Index 001 Part.-No. 00163446
5/11