Electronic Peak Flow Meter (Product Model: DL-DF01) Instruction For Use Effective Date:2022-03-20 Doc. No.
Content Package Contents ................................................................................................... 1 1 Overview ............................................................................................................... 2 2 Product composition ............................................................................................ 4 3 How to use ............................................................................................................
Package Contents No.
Instruction For Use ◆Please read this manual carefully and follow the instructions before you use. ◆Please keep this manual properly so that you can refer to it when you need it in the future. ◆◆We recommend user to use the product exclusively and avoid sharing with others. This device complies with Part 15 of the FCC Rules.
1.2 Function introduction: The electronic peak flow meter is a portable device that can perform lung function tests. It can be used to measure the maximum expiratory flow rate (PEF). The maximum expiratory flow rate (PEF) refers to the highest flow rate during forced exhalation, also known as the highest expiratory flow rate, the maximum expiratory flow rate, and the best expiratory flow rate.
equipment. This equipment should be installed and operated with a minimum distance of 0mm between the radiator and your extremity,60mm for your head. : The patient is an intended operator. The patient can use and maintain the meter and its accessories according to this manual. Put the meter into the specified storage case during transportation. Keep the device out of the reach of children/pets to avoid inhalation or swallowing of small parts.
Blowing Mouthpiece Turbine Main Unit Figure 1. Side view of product “RIGHT” Button “LEFT” Button “ON/OFF” Button Figure 2.
Figure 3. Screen interface picture Materials of components is shown below. Component name Material Housing Acrylonitrile Butadiene Styrene (ABS) Button SEBS Thermoplastic Elastomer Compounding Rubber (SEBS TPR) Screen panel Polycarbonate (PC) Turbine Polycarbonate (PC) Polycarbonate+Acrylonitrile Butadiene Styrene (PC+ABS) Blowing mouthpiece Polypropylene (PP) Meaning of display icon in the screen is shown below.
Male icon Female icon Weight icon Height icon Indicator to traffic colour outside surface of main unit, the position of arrow means the degree of PEF normal or declined. Kilogram (kg) icon PEF% Age Age icon Set Setting icon 3 How to use 3.1. Installation Open the battery cover and insert two AAA batteries into the battery compartment.
Insert the mouthpiece into the turbine according to the direction of arrow marked on the mouthpiece, and press it gently to ensure that there is no air leakage at the connection. Remove the accessories and batteries:remove in reverse order. NOTICE: Install the batteries correctly according to the sign of positive and negative terminals in the battery compartment. 3.2. Boot Long press the “ON/OFF” button for 2 seconds until you hear the “DI” sound, prompting the device has been turned on.
Prompt: 1. After the setting of basic information is completed, the device will calculate an estimated value based on your input parameters, which is used to calculate PEF%. 2. The predicted value is not displayed, but PEF% will be displayed after each measurement is completed. 3.3. Testing Deeply inhale and use your lips to wrap the mouthpiece firmly, and make sure that there is no air leakage between the mouth and the mouthpiece.
recommendation to take appropriate treatment measures, such as increasing the dose of medication or seeking emergency treatment. : Please clean mouthpiece according to the Manual Chapter 4 to prevent from dust because of transport or storage when you use the device for the first time. Do not block the air outlet with your tongue or teeth, otherwise the measurement result will be affected. Mouthoiece must be used by one person,don’t cross-use. 3.4 Remeasure Short press the “ON/OFF” button.
time (display in order: month, day, hour, minute). Waiting for about 2 seconds, the interface will automatically display the PEF value in the corresponding history record number (upper line) and PEF% (lower line). The device can store 500 sets of data. When the memory data exceeds 500 sets of data, the device will automatically overwrite the earlier data in the order of the memory data. 3.6 Data Transmission The device can also be connected to communication devices (such as the phone) through Bluetooth.
2) Click "Bond" and then “This APP needs to use Bluetooth” going to pop up.Click “Allow”,the APP will search automatically. When "Connected to: JDY-08" is displayed, the connection is successful.Bluetooth symbol " then appear on the screen of the device. "will 3) The test can be performed according to 3.3 test steps for measurement, and the measured values will be uploaded to the APP synchronously. 4) Click "Got it" to enter the test interface. 5) Repeat the measurements as per steps 1-4.
Click " " in the test interface and select “Personal information” to modify your name, gender, age, height, weight, etc. 3.6 Shutdown Manual shutdown: long press the “ON/OFF” button for 2 seconds until you hear the "DIDI" sound and the screen interface displays "OFF" flashing twice, and the device will shut down.
4 Cleaning and disinfection 4.1 Cleaning of the turbine NOTICE: Do not immerse the turbine in hot or boiling water. To avoid irreparable damage to the turbine, never try to clean the turbine under a direct jet of water or other liquids. Correct operation of the turbine is guaranteed only if it is "clean" and free of foreign objects that affect its movement. The presence of dust or foreign bodies (such as hairs, sputum etc.
cleaning steps untill it is clean.Then wipe it dry with a clean soft cloth. To disinfect the mouthpiece, wipe the clean mouthpiece with an alcohol cotton pad or a cotton cloth moistened with 75% alcohol for 2 miniutes. Allow to air dry on a towel. Finally store it in a clean, ventilated and dry place. 4.3 Cleaning of the main unit Weekly cleaning is recommended. To clean, wipe the device’s surfaces with a soft damp cloth. Dry with a soft cloth, or allow to air dry.
5.3 Storage Please store the electronic peak flow meter in a clean and dry place under the storage environment condition as specified in the manual. Do not place it under the sun or extreme high or low temperature to avoid severe impact. Otherwise, these may cause the device to fail to work normally or even be damaged. 5.4 Waste disposal The meter, accessories and consumables are to be disposed of according to local law or regulation after they have exceeded their service life.
Screen shows “Err 3” Out of measurement range (PEF Value > 999L/min) Remeasure or replace other equipment to measure The display suddenly disappears. The battery is loose. Reinstall the battery NOTICE: If you take the above measures, the electronic peak flow meter still can not be used normally, it is recommended to contact the manufacturer or dealer.
Resistance to flow <0.
This symbol indicates that waste electrical and electronic equipment must not be disposed of as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment.
IEC 61000-3-2 7 Voltage fluctuations / flicker emissions Not applicable IEC 61000-3-3 NOTICE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. 10.
interruptions and voltage variations on power supply input lines IEC 61000-4-11 for 0.5 cycle at 0°, 45°, 90°, 135°,180°, 225°, 270°, and 315° 0 % UT (100 % dip in UT) for 1 cycle at 0° 70 % UT (30 % dip in UT) for 25/30 cycles at 0° 0 % UT (100 % dip in UT) for 250/300 cycle at 0° Power frequency (50/60 Hz) magnetic field 30 A/m, 50/60Hz 30 A/m, 50/60Hz Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Conducted RF IEC 61000-4-6 3 Vrms 150 kHz to 80 MHz 6 Vrms150 kHz to 80 MHz outside ISM bandsa Portable and mobile RF communications equipment should be used no closer to any part of the Models DL-DF01, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance N/A 10 V/m Radiated RF 10 V/m IEC 61000-4-3 3.5 d P V1 3.5 d P E1 80 MHz to 2.
vicinity of equipment marked with the following symbol: NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Model DL-DF01 as recommended below, according to the maximum output power of the communications equipment. Separation distance according to frequency of transmitter m Rated 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.7 GHz maximum 3 .5 3 .5 7 output of d [ ] P d [ ] P d [ ] P transmitter V1 E1 E1 W 0.01 0.12 0.04 0.07 0.1 0.37 0.12 0.23 1 1.17 0.35 0.7 10 3.7 1.11 2.22 100 11.7 3.5 7.
measurement accuracy lower than stated, and create a potentially unsafe situation. The use of accessories other than those specified, may result in increased emissions or decreased immunity of the equipment or system To maintain compliance with the RF exposure requirement, a separation distance of 20 cm between the device and the human should be maintained. 11 Information of Manufacturer Name: Taian Dalu Medical Instrument Co., Ltd.