Use Instruction

21
interruptions
and
voltage
variations
on power
supply
input lines
IEC
61000-4-11
for 0.5 cycle at 0°,
45°, 90°,
135°,180°, 225°,
270°, and 315°
0 % UT
(100 % dip in UT)
for 1 cycle at 0°
70 % UT
(30 % dip in UT)
for 25/30 cycles at
0°
0 % UT
(100 % dip in UT)
for 250/300 cycle at
0°
Power
frequency
(50/60 Hz)
magnetic field
IEC
61000-4-8
30 A/m, 50/60Hz
30 A/m, 50/60Hz
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: UT is the a. c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The DL-DF01 is intended for use in the electromagnetic environment specified
below. The customer or the user of the DL-DF01 should assure that it is used in
such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance