User Manual

How to do a control solution test?
Step 1- Insert a test strip and check if the code number shown on LCD display
is the same as the code on the strip bottle.
Step 2- Press the M Button. The symbol "CtL" will appear on the display.
When you do this, control solution readings will not be stored in the memory.
Step 3- Get a drop of control solution:
1. Shake the control solution vial well.
2. Remove the cap. Squeeze the vial and
discard the first drop with a clean tissue to
ensure an accurate result.
3. Squeeze the vial again to get a hanging
drop and place the drop on your fingertip.
(Fig.13)
Step 4- Apply the drop to the absorbent hole of the test strip
until the drop is drawn into the test strip, and the confirmation
window is filled. (Fig. 14)
Step 5- The monitor will begin countdown. After 10 seconds, the control
solution test result will appear. The result should fall within the range printed
on the test strip vial.