CALIBRATION
When using the system for the first time, or before using a new vial of test strips, you need to do
calibration, which can be easily done by inserting the code strip into the meter. It is important to
make sure that the code displayed on the LCD is the same as the code printed on the test strip vial
before you proceed. ONLY TEST BLOOD GLUCOSE WHEN THE CODES ARE
IDENTICAL, SO THE TEST RESULTS WILL BE ACCURATE. Please calibrate the meter
according to the following steps.
STEP 1. Enter the Code Mode
Start with the meter turned off. Insert
a test strip to turn on the meter. “CH”
and “ ” will appear. Next, the ambient
temperature and then the code number will appear.
STEP 2 Match the Code Numbers.
Compare the code number on the
meter display with the code number
on the test strip vial. If the two code
numbers match, you may begin testing.
If not, follow Step 3.