Response : Write Multiple Registers
Function code 1 byte 0x10
Register address 2 bytes
Register count 2 bytes
Otherwise an error response is send.
6.5 Calibrate device
This command calibrates the device for external circuits. Analog channels 0 and
1 have to be set to the following levels during calibration.
Channel 0 must be connected to analog ground.
Channel 1 must be connected to the highest expected valid level.
Note : The device is already calibrated at delivery.
Command/Response : Calibrate Device
Function code 1 byte 0x64
If the function succeeds the response is equal to the received command
otherwise an error response is send.
6.6 Save calibration data
Calibration data is loaded at system reset. If another calibration data is needed
by default, use this command after device calibration (see above). A key is
needed to unlock this operation.
Since firmeware version 1.09 this function also saves the baud rate setting to
provide to boot with a customized baud rate.
Note : Saving calibration data will overwrite factory settings! Do not use this
function after having called „Calibrate device“ accidentally.
Command/Response : Save Calibration Data
Function code 1 byte 0x65
Key 2 bytes 0xABCD
If the function succeeds the response is equal to the received command
otherwise an error response is send.