Instructions

106
VOLTage:PROTection:LOW
Sets the under voltage protection (UVP) level of the power supply.
SCPI Command Syntax [SOURce]:VOLTage:PROTection:LOW <NRf+>
GEN Command Syntax UVL < NRf+> , UVP < NRf+>
Default Suffix V
*RST Value 0
Examples VOLT:PROT:LOW 2.5 VOLT:PROT:LOW MAX
Query Syntax
[SOURce]:VOLTage:PROTection:LOW [:LEVel]?
VOLT:PROT:LOW? MIN; VOLT:PROT:LOW? MAX
Returned Parameters
<NR3> VOLT:PROT:LOW? returns presently programmed UVP level.
VOLT:PROT:LOW? MAX and VOLT:PROT:LOW? MIN returns the maximum
and minimum programmable UVP levels.
NOTE:
VOLT:PROT:LOW MAX set UVP level maximum 5% below voltage value
VOLTage:TRIGger
Programs the pending triggered voltage level of the power supply. The pending triggered voltage
level is a stored value that is transferred to the output terminals when a trigger occurs.
SCPI Command Syntax [SOURce]:VOLTage[:LEVel]:TRIGger [:AMPLitude] <NRf+>
GEN Command Syntax None
Default Suffix V
Examples VOLT:TRIG 1200 MV VOLT:LEV:TRIG 1.2
Query Syntax [SOURce]:VOLTage[:LEVel]:TRIGger [:AMPLitude]?
Returned Parameters
<NR3> VOLT:TRIG? returns the presently programmed voltage level.
If the VOLT:TRIG level is not programmed, the default value is 0V.
7.12.4 Current Subsystem
This subsystem programs the output current of the power supply.
CURRent
Sets the output current value in Amperes. The range of current values are described in Tables 7-6,
7-7 and 7-8. The maximum number of characters is 12.
SCPI Command Syntax [SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude] <NRf+>
GEN Command Syntax PC <NRf+> PC?
Default Suffix A
*RST Value 0
Examples CURR 500 MA CURR:LEV .5
Query Syntax
[SOURce]:CURRent[:LEVel][:IMMediate][:AMPLitude]?
[SOURce]:CURRent? MAX [SOURce]:CURRent? MIN
Returned Parameters
<NR3> CURR? returns the present programmed current level.
CURR? MAX and CURR? MIN returns the maximum and minimum programmable
current levels.