Instructions

109
7.12.7 INITiate Subsystem
INITiate
Enables the trigger subsystem. If a trigger circuit is not enabled, all trigger commands are ignored.
SCPI Command Syntax INITiate[:IMMediate]
GEN Command Syntax None
Parameters None
Examples INIT:IMM
Query Syntax None
INITiate:CONTinuous
INIT:CONT 0 - Enables the trigger subsystem only for a single trigger action. The subsystem
must be enabled prior to each subsequent trigger action.
INIT:CONT 1 - Trigger system is continuously enabled and INIT is redundant.
SCPI Command Syntax INITiate:CONTinuous <bool>
Parameters 0|OFF 1|ON
Examples INIT:CONT 1 INIT:CONT ON
Query Syntax INITiate:CONTinuous?
Returned Parameters 0 |1
NOTE:
When power supply is in INIT:CONT ON mode, it’s impossible to change any programming parameter.
Display shows ”Err . Send ABORt commands to allow parameter change.
7.12.8 LIST Subsystem
This subsystem receives parameters to create a step waveform of the output voltage or current.
NOTE:
All list subsystem commands (as well as CURR:MODE LIST and VOLT:MODE LIST) are implied ABORt
commands.
LIST:COUNt
Sets the number of times that the list is executed before it is completed. The command accepts
parameters in the range 1 through 9999, but any number greater than 9999 is interpreted as
INFinity. Use INF if you wish to execute a list indefinitely.
SCPI Command Syntax [SOURce]:LIST:COUNt <NR1>
Parameters 1 to 9999 | INFinity
*RST 1
Examples LIST:COUN 3 LIST:COUN INF
Query Syntax [SOURce]:LIST:COUNt? (if count>9999 the response is INF)
Returned Parameters <NR1>