Recording Equipment User Manual

Analog Data
Digital Data
(Sencitivity, Calibration Data,
Manufacturer Code,
Serial Number, etc.)
LX
IEEE 1951.4 Mixed-Mode Transducer
IEEE1394/Ethernet
Transducer
TEDS
Memory
NETWORK
GPS
LX
LX
LX
Router
Switching Hub
Ethernet
10BASE-T/
100BASE-TX
Etherne
Multiple LXs
Internet/WAN
LX Series
Recording Unit
PRINTED IN JAPAN 1202H05 R-004
Main Units
LX-10
I/O Amp Slots 2 (Eight channels per slot)
Recording Devices Choice of Memory only, Memory + MO disk, or
Memory + PC card drive (Specify one when you
order.)
Interface Choice of IEEE 1394 or 10BASE-T/100BASE-TX
Ethernet (Specify when you order.)
Front Panel Control Keys REC, FWD, STOP, PAUSE, EVENT, P.LOCK
Digital Control Connector Contact input: REC FWD, REC, FWD, STOP,
PAUSE, EVENT,P.LOCK, and ±30-second
internal clock adjustment Status output: REC,
FWD, STOP, PAUSE, EVENT, and P.LOCK
Input: L level: 0.4 V or less, H level: Open or 2 V
or more, Pulse width: 100 ms or more Output:
Open drain, maximum sync current: 8 mA Connector:
Angled half pitch 36- pin (DHA-RC36-R1xxx series)
Monitor Channel 1 (analog output)
Microphone Jack 1
Speaker and Earphone Jack
1 each
Sampling Frequencies 96, 48, 24, 12, 6, 3, 1.5 kHz
(Common to each channel)
Frequency Bandwidths DC to 40, 20, 10, 5, 2.5, 1.25, 0.625 kHz
Max. Recording Rate With real-time transfer to memory or PC: 1.536
MB/s (by IEEE1394)/0.768MB/s (by 100BASE-TX)
(Example: 40 kHz bandwidth
×8 channels in
1,536 MB/s)
With an MO disk/a PC card: 0.768 MB/s
(Example: 20 kHz bandwidth
×8 channels)
Time Precision ±1 ppm (at 25°C)
Internal Clock Correction ±30 seconds' adjustment
Power Supply/Consumption
11 to 30 V DC, approx. 40 W for an 8-channel
I/O configuration (AR-LXPA1 and AR-LXAO)
with media recording
DC power supply cable and AC adapter are supplied.
Safety Standards CE, VCCI
Vibration Conforms to MIL-STD-810E Figure 514.4-1,2,3
for the models with memory only, memory + PC
card drive, Memory with an MO drive only when
an MO disk is not inserted.
Shock Conforms to MIL-STD-810E Method 516.3 20 G
11 ms Half Sine (using PC Memory Card).
External Dimensions Approx. 300W
×65H×200D mm
Approx. 12W
×2.5H×8D in
(excluding protruding parts)
Weight Approx. 3.8 kg Approx. 8.5 lb (for an 8-channel
I/O configuration using an MO disk/a PC card)
Temperature and Humidity PC card model and Memory model: 0 to 40°C
32 to 104°F at 10 to 85 % RH (operation)
MO model: 5 to 35°C 40 to 95°F at 10 to 85 % RH
(operation)
Software The LX Navi software, provided as standard, can
set the measurement parameters, monitor, and
record data to the PC. It has the following system
requirements. OS: Windows 98SE/Me/2000/XP
CPU: Pentium III 600 MHz or faster RAM: 128 MB
or more Free Hard Disk Space: 2 GB or more
LX-20
Sampling Frequencies - 96, 48, 24, 12, 6, 3, 1.5 kHz- 102.4, 51.2,
25.6, 12.8, 5.12, 2.56, 1.28 kHz
- 65.536, 32.768, 16.384, 8.192, 4.096, 2.048, 1.024 kHz
- 100, 50, 20, 10, 5, 2, 1 kHz
(Common to each channel)
Frequency Bandwidths DC to the sampling frequency (listed above) / 2.4
Tachometer Pulse Inputs
Num. of Input Channels 2
×16-bit channels (ON/OFF in 2-channel units),
or 1
×32-bit channel
Both use the lower 1 bit for tachometer pulse timing.
Input Format Threshold level switched over among +0.5, 1,
2.5, 5, 10, and 20 V
Max. allowable input voltage is ±50 V
Input Connector
2×BNCs
Sampling Frequency
Depends on the main unit setting.
Frequency Division Ratio Setting
1 to 255
Moving Average Measurement
1 to 16 (Cannot be used simultaneously with
generator output.)
Measurement Modes
Pulse count mode (Count of number of pulses
within the gate time; count of the total number from
the start to the stop.)
Cycle count mode
Frequency measurement mode (Cannot be used
simultaneously with generator output.)
RPM mode (Cannot be used simultaneously with
generator output.)
Generator Output
Num. of Output Channel
1 channel
Output Level
±1 to 5 Vp (variable in 0.1 V steps, same as
monitor output)
Output Connector
BNC (Monitor output)
Output Mode
Sine wave 1 to fs/2.4 (to 1 digit below the decimal
point)
Sweep Sine wave, start/stop frequency, linear
and log settings.
Pulse, pulse width, and interval time settings
(There are restrictions depending on the
sampling frequency.)
Pink noise
White noise
Power Supply/Consumption
11 to 30 V DC, approx. 45 W for an 8-channel I/O
configuration (AR-LXPA1 and AR-LXAO) with media
recording
DC power supply cable and AC adapter are supplied.
Weight Approx.4.1 kg Approx.9 lb (for an 8-channel I/O
configuration using an MO disk/a PC card)
* Other specifications follow the LX-10 specifications.
Battery Enclosure(BU-80)
Num. of Internal Battery Packs
3 (A battery pack is described below)
External Dimensions Approx. 300W
×27.5H×200D mm
Approx. 12W
×1H×8D in
(excluding protruding parts)
Weight Approx. 1.5 kg Approx. 3 lb
(excluding the battery pack and mounting brackets)
Battery Pack (HP-30L )
Supplied Voltage 13.2 V
Capacity 3.3 Ah
Weight Approx. 700 g Approx. 1.5 lb
Size NP1 type
Battery Type Ni-MH
Remote Control Unit (ER-LXRC)
LCD Reflective type, 320
×160 pixels
Functions Bar meter display, mainunit control (setting,
recording, reproducing) Event, Microphone input
External Dimensions Approx. 170W
×30H×100D mm
Approx. 6.5W
×1.2H×4D in
(excluding protruding parts)
Cable length 5 m 16 ft
Weight Approx. 0.65 kg Approx. 1.4 lb(excluding cable)
Extension I/O Unit for DC Input Amp (AU-LXEPIO)
I/O Amp Slots 2 (Specify the combination of amp cards to be
built in when you order.)
External Dimensions Approx. 300W
×30H×200D mm
Approx. 12W
×1.2H×8D in
(excluding protruding parts)
Weight Approx. 1.2 kg Approx. 2.5 lb
(excluding the amp card)
Extension I/O Unit for the Constant Current
Excitation Type Accelerometer Input Amp
(AU-LXEPIOP)
I/O Amp Slots 2 (Eight channels per slot)
(Specify the combination of amp cards to be built
in when you order.)
Power Supply 11 to 30 V DC, DC power supply cable and AC
adapter are supplied.
External Dimensions Approx. 300W
×50H×200D mm
Approx. 12W
×2H×8D in
(excluding protruding parts)
Weight Approx. 2 kg Approx. 1 lb (excluding the amp card)
Amp Cards
DC Input Amp (AR-LXDC2)
Num. of Channels 8
Input Format Unbalanced
Input Coupling DC
Low-speed Sampling Fixes the sampling at 2 kHz, with DSP-based
down sampling.
Input Impedance 100 kΩ
Input Range ±0.5, 1, 2, 5, 10, 20, 50 Vp (overrange ±127%)
Absolute Max. Input Voltage
±100 V
Filter Joint use of both a digital filter (Decimation LPF)
and an analog filter (2nd Order Butterworth)
Frequency Characteristics Tolerance of ±0.5 dB for DC to each bandwidth
of the main unit
Num. of Quantizing Bits 24 bits, MSB16 bits used for data out of 24 bits
Conversion Method Simultaneous sampling, 128-times oversampling
(delta sigma) method; 64-times
oversampling at 96 kHz sampling
Sampling Frequencies See the sampling frequencies of the main unit.
Power Consumption Approx. 5 W
Constant Current Excitation Type Accelerometer
Input Amp
(AR-LXPA1)
Num. of Channels 8 differential inputs
Input Format Balanced and unbalanced
Input Coupling Balanced AC coupling, balanced DC coupling,
unbalanced DC coupling
Input Impedance 100 kΩ
Input Range ±0.01, 0.0316, 0.1, 0.316, 1.0, 3.16, 10 Vp
(overrange to ±127%)
Absolute Max. Input Voltage
±50 V
Filter Joint use of both a digital filter and an analog filter
(2nd Order Butterworth)
Frequency Characteristics
±0.5dB. In AC mode: 1 Hz to the bandwidth of
the main unit (-3 dB at 0.1 Hz). In DC mode: DC
to the bandwidth of the main unit.
Num. of Quantizing Bits 24 bits, MSB16 bits used for data out of 24 bits
Conversion Method Simultaneous sampling, 128-times oversampling
method; 64-times oversampling at
96 kHz sampling
Sampling Frequencies See the sampling frequencies of the main unit.
Power Supply to Sensors 28V DC, 4 mA
Power Consumption Approx. 7 W
Constant Current Excitation Type Accelerometer
Input Amp
(AR-LXPA2)
Input Range ±0.0316, 0.1, 0.316, 1.0, 3.16, 10, 50 Vp
(overrange to ±127%)
Absolute Max. Input Voltage
±50 V in the ±10 Vp range, ±100 V in the
±50 Vp range
Weighting A, C, Flat
High Pass Filter 10 Hz, 20 Hz switchable
Power Supply to Sensors 28V DC/4 mA or 24V DC/4 mA switchable at
hardware
* Other specifications follow the AR-LXPA1 specifications.
Analog Output Amp (AR-LXAO)
Num. of Channels 8
Output Format Unbalanced
Output Coupling DC coupling
Output Level ±1 to 5 V (variable in 0.1 V steps)
Frequency Bandwidths DC to the frequency bandwidth of the main unit
Filter Joint use of both a digital filter (Decimation LPF)
and an analog filter(2nd Order Butterworth)
Num. of Quantizing Bits 24 bits, MSB16 bits used for data out of 24 bits
Conversion Method 128-times oversampling
method; 64-times
oversampling at 96 kHz sampling
Output Connector BNC
Power Consumption Approx. 5 W
Overall Accuracy
When Using AR-LXPA1 and AR-LXAO
Range accuracy ±2% or less
Linearity ±0.1% or less
Distortion factor ±0.05% or less (for bandwidth of 20 kHz or
more),±0.2% or less (for bandwidth of 10 kHz or
less)
Drift ±0.1% or less (from 10 minutes after the power
has been turned on and when recording to one
media)
Inter-channel phase difference
Typical 0.5° (in the same range and with a
bandwidth of 20 kHz or less), or 1° or less (in the
same range and with a bandwidth of 40 kHz)
Recording and playback
performance that exceeds
DAT Data Recorders
Connecting with Applications
The LX Series provides a diverse set of connections to popular data analysis applications, which enables
efficiency in all steps of the process: from experiment setup to data recording and processing.
Connecting to Sensors
Using the constant current excitation type accelerometer input amp card, you can directly connect to ICP
®
accelerometers and ICP
®
microphones. You can also read values such as sensitivity and calibration values
for connected transducers. For experiment setup, these features reduce the need for additional equipment
and can automate coefficient settings, thereby eliminating setup errors, and greatly reducing the effort
required for experiment setup and data processing.
Connecting to Existing Analysis Systems
Some existing data recorders have the ability to
perform analog playback and change the playback
speed (matching the analysis system bandwidth
and playing back fast phenomena at a slower
speed, or playing back slow phenomena at a faster
speed, within the same sampling rate series)
The LX Series can replace a traditional data recor-
der even in an analysis system that requires exist-
ing analog data to be played back. At a PC with
the optional LX View program, you can extract
only a necessary portion from an originally recor-
ded file to create a new data file. Then you can
copy the new data file to another media to per-
form analog playback in the LX main unit. This
enables data to be investigated more efficiently.
Connecting to Data Processing Software
The format for recorded data is TAFFmat, the file
format now used by TEAC digital Data Recorders.
Data files in the TAFFmat format can be read by
the LX View and by other companies' analytical
software that supports TAFFmat. Also, in a config-
uration in which a PC is connected and a LX Ser-
ies recorder is used as the front-end, the TAFFmat
format files can be used with software providing
real-time recording and analysis functions. We
also pro-vide the file converter software from
TAFFmat to the popular analyzing data file formats,
such as ASCII, UFF, ATI, and so on. TEAC provides
system integrators with a Windows DLL for direct
control of LX Series recorders. Contact TEAC for
details.
Connecting to Network
Using the 10BASE-T/100BASE-TX Ethernet interface, you can build distributed multi-channel data record-
ing systems. Synchronizing pulses generated by, for example, a GPS signal can synchronize the internal
clocks by using +/-30 seconds' adjustment function of the LX and you can control the recording of
multiple devices connected to the network. The data collected by each LX main unit can be collected
over the network, which reduces the effort required to physically collect media from different locations.
Specifications
S/N (Crosstalk), measured at the Analog Output:
When Using AR-LXDC2 and AR-LXAO
S/N 78 dB (within the bandwidth)
Crosstalk
- 78 dB (below 20 kHz within the bandwidth)
or 75 dB (40 kHz within the bandwidth)
(except for low-sped sampling)
Note: The overall accuracy specified here is the accuracy of the analog output
when using an input amp and a DC output amp, and may differ depending on
the frequency characteristics and the input range of the input amp.
TEAC INSTRUMENTS CORPORATION
83 Imaikami-cho, Nakahara-ku, Kawasaki-shi, Kanagawa 211-0067, Japan Phone: +81-44-711-5221 Fax: +81-44-711-5240 http://www.tic.teac.co.jp/
TEAC AMERICA INC. / Information Product Division
7733 Telegraph Road, Montebello, CA 90640 U.S.A.
West: Phone: +1-323-726-0303 Ext.573 Fax: +1-323-727-4877 East: Phone: +1-978-468-4135 Fax: +1-978-468-4171
http://www/teac-recorders.com/drindex.html
TEAC DEUTSCHLAND GmbH / ICP Division
Barnstrasse 12, D65205 Wiesbaden-Erbenheim, Germany Phone: +49-611-7158417 Fax: +49-611-7158392 http://www.teac.de/ipd/
Precautions: To ensure safe operation, read the instruction manual before use.
The trademarks in this brochure are owned by their respective holders.
Specifications are subject to change without notice.
Copyright © 2002, TEAC Instruments Corporation All rights reserved.
The new LX Series data re-
cording system from TEAC
has been designed for fast
set-up, reliable recording, and
versatile playback in your
data recording applications in
the laboratory and in the field.
Following-on from the conve-
nience of DAT recorders, the
LX Series en-ables recording
in wider bandwidths. The ser-
ies has strengthened connec-
tivity with sensors, PCs and
analysis applications; achieves
fast data processing; and is
designed to lower data re-
cording costs.
TEAC Instrumentation Recording Systems
Note : Analog voltage output is not supported.
80dB
80dB
80dB
80dB
80dB
60dB
60dB
40dB
No filter
No filter
No filter
Sampling
Frequency
Frequency
Bandwidth
1/2 fs Attenuation
1kHz
500Hz
200Hz
100Hz
50Hz
20Hz
10Hz
5Hz
1Hz
1/30Hz
1/60Hz
DC to 400Hz
DC to 200Hz
DC to 80Hz
DC to 40Hz
DC to 20Hz
DC to 8Hz
DC to 4Hz
DC to 2Hz
DC to 0.4Hz
DC to 0.013Hz
DC to 0.006Hz
Input Rage At 20 kHz or less
At 40 kHz
±0.01V
±0.0316V
±0.1V
±0.316V
±1.0V
±3.16V
±10V
64 dB (- 64 dB)
74 dB (- 71 dB)
83 dB (- 77 dB)
87 dB (- 77 dB)
87 dB (- 77 dB)
87 dB (- 77 dB)
87 dB (- 77 dB)
60 dB (- 60 dB)
69 dB (- 69 dB)
76 dB (- 73 dB)
77 dB (- 73 dB)
77 dB (- 73 dB)
77 dB (- 73 dB)
77 dB (- 73 dB)

Summary of content (2 pages)