Operating Manual
13
NelumBox® is equipped with an electronic access control using RFID
technology (optional). If the electronic lock is activated, the door can
only be unlocked by RFID tokens/cards or via NFC.
NelumBox® can only be opened when switched on (exception: emer-
gency release; see chapter 6.4.1). To unlock NelumBox®, hold the RFID
token/card to designated area at the front of the device. Successfully
unlocking NelumBox® is confirmed by a green light on the buttons and
via the display.
6.6 Opening / closing NelumBox®
If unlocking is prevented, do not attempt to open the door by force. This
can lead to damage to the device.
• The lock is unlocked by pulling either the upper or the lower
button and by pulling the door. Prevented access is confirmed
by the buttons glowing in red and on the display.
• If the electronic lock of NelumBox® is not activated, simply
open the device by pulling either the upper or the lower
button and by pulling the door.
• If the unit is opened for more than 22 seconds, a warning
signal sounds and the door lights begin to shine in yellow.
The signal warns to quickly close Nelumbox®’s door to avoid
exceeding the set temperature. If the alarm message is
ignored, the signal increases in intensity. The display also
prompts the user to close the door.
• Close the door by applying pressure to the front of NelumBox®.
Press on the area between the display and the hinged handle
or on the LED to close the door.
• Do not press on the display and do not push any button while
closing the door.
• After successfully closing the door, the audible signal, the
lights and the warning on the display disappear.
Warning!
Warning!