Specifications
This appendix details the Specifications of the GB1400 Generator (TX) and GB1400 Analyzer (RX).
Note: The same term can be expressed three different ways.
CLOCK
= CLOCK BAR = NOT CLOCK
DATA
= DATA BAR = NOT DATA
GB1400 Generator (TX)
Internal Clock Source
Frequency Range 1 MHz to 1400 MHz
Step Size Range 0.01,0.1,1,10,100,1000 MHz
Resolution 1 kHz
Accuracy 10 ppm (within calibration interval)
Frequency Memory 10 frequencies
External Clock Source
Frequency Range 1 MHz to 1400 MHz
Burst Mode Option 150 kHz to 1400 MHz
Input Level 0.5V to 2.0Vp-p
Impedance 50 Ohm, AC coupled
(with Burst Mode option, 50 Ohm ECL)
Connector SMA